Accuray Cyberknife Xchange Robotic Collimator Changer
Accuray is recalling the Xchange Robotic Collimator Changer for the Cyberknife system due to a risk of the collimator falling from its housing.
Immediately cease using the in-room control pendant to initiate motion if a system interlock has occurred. Contact Accuray Incorporated for instructions on system inspection and software updates to resolve the docking failure.
Recall at a glance
- Quantity recalled
- 292 units
Why it was recalled
Accuray is recalling the Xchange Robotic Collimator Changer used with the Cyberknife Treatment Delivery System. A failure in the docking mechanism can cause the collimator to release and fall from its housing if motion is initiated via the in-room control pendant after a system interlock occurs.
Is mine affected?
This recall covers the product below. If you have it, yours is affected.
Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator
Where it was sold
Distributed nationwide.
Timeline
- Firm initiated recall
- Publicly announced
- Listed on Recall Canary
- Summary updated
Official source
Other Device Defect recalls
Other recent recalls from Accuray Incorporated
Editorial summary by Recall Canary · summary updated 2026-07-02 · primary source: U.S. Food and Drug Administration · severity reflects the agency's published classification