What happened
Baxter Healthcare Corporation recalled Likorall Overhead Lifts because a potential issue involving motor shaft breakage combined with reduced performance of the Single Fault Safety (SFS) system may lead to instances where the lift strap lowers more quickly than expected during use. Patients are at some risk for injury if a significant fall/drop were to occur due to this issue. The company began the recall on September 14, 2026, and the FDA added it to the Enforcement Report on October 13, 2026.
Summary by Recall Canary from the FDA notice.
What to do
We don't have product-specific instructions for this recall. FDA's general advice for medical device recalls:
- A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
- Talk to your health care provider before you stop using it.
Is yours affected?
Compare the details below with your product and check the FDA notice to confirm whether it is affected.
Incomplete list: 40 of 53 UDI-DIs shown.
Likorall Overhead Lifts
Product details from the notice
Likorall Overhead Lifts Product Code: 3121001 Product Description: LIKORALL 200 Product Code: 3122005 Product Description: LIKORALL 242 ES, NATURAL Product Code: 3122005NA Product Description: LIKORALL 242 ES, NATURAL Product Code: 3122006 Product Description: LIKORALL 242 ES, WHITE Product Code: 3122006NA Product Description: LIKORALL 242 ES, WHITE Product Code: 3122007 Product Description: LIKORALL 242 ES R2R, NATURAL Product Code: 3122007NA Product Description: LIKORALL 242 ES R2R, NATURAL Product Code: 3122008 Product Description: LIKORALL 242 ES R2R, WHITE Product Code: 3122008NA Product Description: LIKORALL 242 ES R2R, WHITE Product Code: 3122009 Product Description: LIKORALL 242 S, NATURAL Product Code: 3122009NA Product Description: LIKORALL 242 S, NATURAL Product Code: 3122010 Product Description: LIKORALL 242 S, WHITE Product Code: 3122011 Product Description: LIKORALL 242 S R2R, NATURAL Product Code: 3122011NA Product Description: LIKORALL 242 S R2R, NATURAL Product Code: 3122012 Product Description: LIKORALL 242 S R2R, WHITE Product Code: 3122012NA Product Description: LIKORALL 242 S R2R, WHITE Product Code: 3122501 Product Description: LIKORALL 250 ES, NATURAL Product Code: 3122501NA Product Description: LIKORALL 250 ES, NATURAL Product Code: 3122502 Product Description: LIKORALL 250 ES, WHITE Product Code: 3122502NA Product Description: LIKORALL 250 ES, WHITE Product Code: 3123001 Product Description: LIKORALL 243 ES Product Code: 3123001NA Product Description: LIKORALL 243 ES Product Code: 3123002 Product Description: LIKORALL 243 ES, WHITE Product Code: 3123002NA Product Description: LIKORALL 243 ES, WHITE Product Code: 3124050 Product Description: LIKORALL 250 S, IRC
- UDI-DI
- 00887761981911
- 3122005
- 242
- 00887761979710
- 3122005NA
- 00887761979727
- 3122006
- 00887761979734
- 3122006NA
- 00887761979741
- 3122007
- R2R
- 00887761979758
- 3122007NA
- 00887761979765
- 3122008
- 00887761979772
- 3122008NA
- 00887761979789
- 3122009
- 00887761979796
- 3122009NA
- 00887761979802
- 3122010
- 00887761979819
- 3122011
- 00887761979444
- 3122011NA
- 00887761979451
- 3122012
- 00887761979468
- 3122012NA
- 00887761979475
- 3122501
- 250
- 00887761979482
- 3122501NA
- 00887761979499
- 3122502
- 00887761979505
Details
- Company began the recall
- Event ID
- 99802
- Company
- Baxter Healthcare Corporation
- Hazard
- Device defect
Company began the recall
Added to FDA's Enforcement Report · Classified by FDA