Medical Action Industries Pack Cath BHH
← Back to all recalls
MEDICAL DEVICE RECALLCRITICAL

Medical Action Industries Pack Cath BHH

Medical Action Industries is recalling Pack Cath BHH kits due to a potential syringe adapter disconnection issue.

What to do

Stop using the affected convenience kits immediately and contact the company for further instructions regarding return or replacement.

Posted
Updated
Recall No.
Z-2007-2026
Agency
Firm
Medical Action Industries, Inc. 306
Recall ID
98698
Status
ONGOING

Why it was recalled

Medical Action Industries is recalling the Pack Cath BHH convenience kit because the rotating adapter on the included syringe may unwind during use. This defect can cause a loose connection or full disconnection between the syringe and the manifold, potentially compromising the procedure.

Is mine affected?

Check the model, lot, or serial number on the device or its packaging. Yours is affected if it matches one below.

Medical Action Industries Inc. Pack Cath BHH, REF: BHCA49K, Sterile EO, Rx Only, QTY: 5/Case

UPC
20809160490294
Recall #
Z-2007-2026

Where it was sold

Distributed in Arizona.