StatStrip Glucose Meters

The StatStrip Glucose Hospital Meter System may show a previously scanned barcode.

High risk

There is a reasonable chance that using it could cause serious health problems or death.

FDA Class I recall · How risk is rated

What to do

What happened

Nova Biomedical Corporation, of Waltham, Massachusetts, is recalling certain StatStrip Glucose and B- Ketone meters and StatStrip Glucose Hospital Meter Systems.

A software synchronization/timing issue in the barcode scanning and patient data processing component causes the device to intermittently retain a previously scanned barcode, including patient identification (ID) barcodes and quality control (QC) solution ID barcodes, when a new barcode is scanned, resulting in the prior barcode information remaining displayed on the screen for the current test.

The recall began on July 29, 2026, and was added to the FDA's Enforcement Report on September 11, 2026.

The products were distributed nationwide and in foreign countries including Austria, Belgium, Bolivia, Brazil, Canada, Cayman Islands, Chile, Croatia, Dominican Rep., Finland, France, Germany, Ireland, Israel, Italy, Netherlands, New Zealand, South Korea, Switzerland, Thailand, Turkey, and the United Kingdom.

Summary by Recall Canary from the FDA notice.

What to do

We don't have product-specific instructions for this recall. FDA's general advice for medical device recalls:

  1. A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
  2. Talk to your health care provider before you stop using it.

Is yours affected?

Match your product and all the details in one listed combination.

  1. StatStrip Glucose Hospital Meter System.

    1. Model
      • 63685
      UDI-DI
      • 0038548063685
    2. Model
      • 66739
      UDI-DI
      • 00385480636858
  2. StatStrip Glucose and B- Ketone meter.

    1. Model
      • 63683
      UDI-DI
      • 00385480664295
    2. Model
      • 63910
      UDI-DI
      • 00385480639101
    3. Model
      • 66429
      UDI-DI
      • 00385480636834

Details

Company began the recall
Sold in

Worldwide distribution: US (nationwide) and OUS (foreign) countries of: Austria, Belgium, Bolivia, Brazil, Canada, Cayman Islands, Chile, Croatia, Dominican Rep., Finland, France, Germany, Ireland, Israel, Italy, Netherlands, New Zealand, South Korea, Switzerland, Thailand, Turkey, United Kingdom.

Event ID
99546
Recall numbers
Z-3128-2026 to Z-3132-2026
Company
Nova Biomedical CorporationWaltham, Massachusetts
Recall type
Voluntary: Firm initiated
Customers told by
E-Mail
  1. Company began the recall

  2. Added to FDA's Enforcement Report · Classified by FDA

  3. Listed on Recall Canary

  4. In FDA's weekly enforcement report

  5. Summary updated

Summary by Recall Canary, written with AI · summary updated Oct 2, 2026 · primary source: U.S. Food and Drug Administration