Philips MultiDiagnost Eleva
Philips recalls MultiDiagnost Eleva systems over hand switch button that may not fully release, risking radiation exposure.
Contact Philips Medical Systems for inspection and repair of the hand switch button. Do not use the system until the issue is resolved.
Why it was recalled
Philips Medical Systems is recalling MultiDiagnost Eleva X-ray systems due to a hand switch button that may not fully release, potentially causing loss of imaging functionality or unintended radiation exposure and additional contrast injection.
Is mine affected?
Check the model, lot, or serial number on the device or its packaging. Yours is affected if it matches one below.
| Product | Recall # |
|---|---|
| Philips MultiDiagnost Eleva with the following Model Numbers | Z-2411-2026 |
| Philips Allura Xper with the following Model Numbers: Model # 722001 for Allura Xper FD10C; Model # 722002 for Allura Xper FD10F; Model # 722003 for Allura Xper FD10; Model # 722005 for Allura Xper FD10/10; Model # 722006 for Allura Xper FD20; Model # 722008 for Allura Xper FD20 Biplane; Model # 722010 for Allura Xper FD10; Model # 722011 for Allura Xper FD10/10; Model # 722012 for Allura Xper FD20; Model # 722013 for Allura Xper FD20 Biplane; Model # 722015 for Allura Xper FD20 OR Table; Model | Z-2412-2026 |
| Philips Allura Xper with the following Model Numbers: Model # 722001 for Allura Xper FD10C; Model # 722002 for Allura Xper FD10F; Model # 722003 for Allura Xper FD10; Model # 722005 for Allura Xper FD10/10; Model # 722006 for Allura Xper FD20; Model # 722008 for Allura Xper FD20 Biplane; Model # 722010 for Allura Xper FD10; Model # 722011 for Allura Xper FD10/10; Model # 722012 for Allura Xper FD20; Model # 722013 for Allura Xper FD20 Biplane; Model # 722015 for Allura Xper FD20 OR Table; Model # 722019 for Allura Xper FD10/10 OR Table; Model # 722020 for Allura Xper FD20 Biplane OR Table; Model # 722022 for Allura Xper FD10 OR Table; Model # 722023 for Allura Xper FD20 OR Table; Model # 722025 for Allura Xper FD20 Biplane OR Table; Model # 722026 for Allura Xper FD10; Model # 722027 for Allura Xper FD10/10; Model # 722028 for Allura Xper FD20; Model # 722029 for Allura Xper FD20/10; Model #722033 for Allura Xper FD10 OR Table; Model #722035 for Allura Xper FD20 OR Table; Model #722038 for Allura Xper FD20/20; Model #722039 for Allura Xper FD20/20 OR Table; Model #722058 for Allura Xper FD20/15; Model #722059 for Allura Xper FD20/15 OR Table. | – |
Official source
Other recent recalls from PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.