Diagnostica Stago, Inc.
Diagnostica Stago, Inc.: 4 recalls listed since January 2026 across FDA.
By year
- 20264
By agency
- FDA4
By category
- Medical Device4
All recalls (4)
STA Staclot Protein S
STA Staclot Protein S has a positive drift in QC values within its stated 4-hour onboard stability period.
STA Liatest D-Di
After receiving customer complaints, investigations confirmed the presence of a positive bias in D-Dimer measurements throughout the entire analytical range.
STA Liatest Free Protein S
The STA Liatest Free Protein S kit may show out-of-range results and underestimate free protein S levels in normal plasma.
Diagnostica Stago Asserachrom HPIA Kits
Asserachrom HPIA can produce negative results due to a defect affecting every well in the impacted strip.