FDA recall, , High risk
Abbott Diabetes Care FreeStyle Libre 3, Lingo and Libre Apps
Abbott FreeStyle Libre 3, Lingo and Libre Apps have a security vulnerability and may lose sensor connection without reconnecting.
What to do
Check yoursIf a code on your device is listed below:
- Check the lot against the recall notice and ask your provider or supplier before you stop using it.
Typical advice for a recalled medical device; the FDA notice gives no product-specific instructions.
What happened
Abbott Diabetes Care, Inc., of Alameda, California, is recalling three software apps used with its diabetes glucose monitoring systems. The company started the recall on its own. The recall covers two problems. The first is a security vulnerability. The second is extended signal loss, in which a sensor loses connection with the app and does not reconnect. The recall covers 1,836,506 app downloads in total, including 1,382,849 in the United States and 453,657 outside the United States.
The company began the recall on October 15, 2026, and told customers by letter. The affected apps are the FreeStyle Libre 3 App Software, versions 3.2.0 through 3.6.8; the Libre App, versions 1.1.0 through 1.4.0; and the Lingo Biosensor App, versions 1.6.0 through 1.12.2 and 1.9.0 through 1.12.2. The apps were distributed nationwide in the United States and in other countries, including Canada, Australia, and Germany. The FDA classified the recall on November 25, 2026.
Is yours affected?
Compare the details below with your product and check the FDA notice to confirm whether it is affected.
A UDI-DI is the device identifier printed with the barcode on the device label.
Libre App
Product details from the notice
Libre App Model/Catalog Number: 78745-01 UDI-DI code: 00357599840006 Software versions: Versions 1.1.0 through 1.4.0 Model/Catalog Number: 78746-01 UDI-DI code: 00357599850005 Software versions: Versions 1.1.0 through 1.4.0 The FreeStyle Libre 3 Continuous Glucose Monitoring System
- UDI-DI
- 00357599840006
- 00357599841003
Abbott Diabetes Care FreeStyle Libre 3, Lingo and Libre Apps
Product details from the notice
Model/Catalog Number: 72077-01 / Model/Catalog Number: 72075-01
- Recall number
- Z-0536-2027
- Product
- Model/Catalog Number: 72077-01
- UDI-DI
- 00357599817008
- Product
- Model/Catalog Number: 72075-01
- UDI-DI
- 00357599816001
Abbott Diabetes Care FreeStyle Libre 3, Lingo and Libre Apps
Product details from the notice
Model/Catalog Number: 78783-01 / Model/Catalog Number: 78782-01
- Recall number
- Z-0537-2027
- Product
- Model/Catalog Number: 78783-01
- UDI-DI
- 00357599849009
- Product
- Model/Catalog Number: 78782-01
- UDI-DI
- 00357599850005
Recall details
- Posted by FDA
- Recall began
- Status
- Open
- Weekly archive
- Listed the week of Nov 23, 2026
- Sold in
Distributed nationwide and in Andorra, Austria, Belgium, Switzerland, Germany, Denmark, Spain, Finland, France, United Kingdom, Italy, Lichtenstein, Luxembourg, Netherlands, Norway, San Marino, Sweden, Qatar, Saudi Arabia, Australia, New Zealand and Canada.
- Recalled by
- Abbott Diabetes Care, Inc.Alameda, California
- Hazard
- Software defect
- Risk
High risk
There is a reasonable chance that using it could cause serious health problems or death.
FDA Class I recall · How risk is rated
- FDA event number
- 100151
- Recall numbers
- Z-0536-2027 to Z-0538-2027
- Recall type
- Started by the company
- How the company first told customers
- Letter
All recalls from Abbott Diabetes Care, Inc.
Timeline
FDA's Enforcement Report is its public list of recalls; FDA posts a recall by adding it there.
Abbott Diabetes Care began the recall
FDA posted the recall to its Enforcement Report and rated it Class I (high risk)