i-STAT Blood Gas Cartridges
← Back to all recalls
MEDICAL DEVICE RECALLMODERATE

i-STAT Blood Gas Cartridges

Abbott is recalling i-STAT EG7+, EG6+, and G3+ cartridges due to inaccurate blood gas results that may lead to incorrect clinical decisions.

What to do

Healthcare facilities should immediately review their inventory for the affected cartridge lots, discontinue use of the recalled products, and follow the instructions provided in the manufacturer's Urgent Medical Device Correction notice to return or dispose of the items.

Posted
Updated
Recall No.
Z-2001-2026
Agency
Firm
Abbott Point Of Care Inc.
Recall ID
98635
Status
ONGOING

Why it was recalled

Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.

How to identify it

UPC 00054749005673

i-STAT G3+ cartridge;

Where it was sold

Distributed in US.