Skip to main content
Recall Canary
i-STAT EG7+ Cartridge
← Back to all recalls
MEDICAL DEVICE RECALLMODERATE

i-STAT EG7+ Cartridge

Abbott recalls i-STAT EG7+ cartridges (03P76-25) over inaccurate PCO2/pH results.

What to do

Check your inventory for i-STAT EG7+ cartridges with list number 03P76-25 and discontinue use of affected lots. Contact Abbott Point of Care for instructions on returning or disposing of affected cartridges and to obtain replacement product.

Recalled
Listed
Updated
Recall No.
Z-1999-2026
Agency
FDA
Firm
Abbott Point Of Care Inc.
Recall ID
98635
Status
ONGOING

Why it was recalled

Abbott Point of Care is recalling i-STAT EG7+ cartridges (list number 03P76-25) because approximately 7.6% of cartridges from specific lots may report higher than expected PCO2 and lower than expected pH results, which could lead to unnecessary or harmful interventions for patients.

Is mine affected?

Check the model, lot, or serial number on the device or its packaging. Yours is affected if it matches one below.

ProductRecall #UDI-DI
i-STAT G3+ cartridge; List Number: 03P78-26;Z-2001-202600054749005673
i-STAT EG7+ cartridge; List Number: 03P76-25;Z-1999-202600054749000043
i-STAT EG6+ cartridge; List Number: 03P77-25;Z-2000-202600054749000050

Where it was sold

Distributed nationwide.

Timeline

  1. Firm initiated recall
  2. Publicly announced
  3. Listed on Recall Canary
  4. Summary updated
← Back to all recalls

Editorial summary by Recall Canary · summary updated 2026-09-08 · primary source: U.S. Food and Drug Administration · severity reflects the agency's published classification