Skip to main content

FDA recall, , Moderate risk

Abiomed Impella Controllers

The Automated Impella Controller has defects in firmware and software that can affect pressure signal accuracy and cause unreliable signals.

What happened

Abiomed, Inc., of Danvers, Massachusetts, recalled 3,322 Automated Impella Controllers (AIC) nationwide and in Austria, Belgium, Canada, France, Germany, Hong Kong, Italy, Luxembourg, Poland, Spain, Switzerland, and the United Kingdom.

The recall, begun February 19, 2019, is a retrospective report of issues remedied in 2019 with firmware and software patches: defects in optical bench (OPB) firmware V2.19, which affected the accuracy and stability of the pressure signal the AIC system performance depends on, and software defects leading to unreliable or missing placement signal, unresponsive start Impella soft key during disconnect/reconnect, and false Position in Ventricle alarm in certain AICs.

The FDA added the recall to its Enforcement Report on September 21, 2026, and classified it the same day. The products include the Impella Controller, Packaged, CA, EU, UK, and US, and the Impella Optical Controller, Packaged, EU.

What to do

We don't have product-specific instructions for this recall. FDA's general advice for medical device recalls:

  1. A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
  2. Talk to your health care provider before you stop using it.

Is yours affected?

Compare the details below with your product and check the FDA notice to confirm whether it is affected.

A UDI-DI is the device identifier printed with the barcode on the device label.

  1. Product details

    Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA;
    Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU;
    Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK;
    Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US;
    Product Code: 0042-0000-US. 5. Impella Optical Controller, Packaged, EU;
    Product Code: 0042-0010-EU. 6. Impella Optical Controller, Packaged, US;
    Product Code: 0042-0010-US. 7. Optical, AIC, Impella Connect, Pkgd, US;
    Product Code: 0042-0040-US.

    UDI-DI
    • 00813502011272
    • 00813502011289
    • 00813502011296
    • 00813502010022
    • 00813502010961
    • 00813502010985
    • 00813502011401

Recall details

Company began the recall
Sold in

Worldwide - US Nationwide and the countries of Austria, Belgium, Canada, France, Germany, Hong Kong, Italy, Luxembourg, Poland, Spain, Switzerland, United Kingdom.

Company
Abiomed, Inc.Danvers, Massachusetts
Risk

Moderate risk

Using it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.

FDA Class II recall · How risk is rated

Event ID
99747
Recall number
Z-3251-2026
Quantity recalled
3,322 units

All recalls from Abiomed, Inc.

Timeline

FDA's Enforcement Report is its public list of recalls; FDA adds a recall when it posts it and publishes the list weekly.

  1. Company began the recall

  2. Added to FDA's Enforcement Report · Classified by FDA

  3. In FDA's weekly enforcement report

  4. Listed on Recall Canary