Skip to main content

FDA recall, , Moderate risk

HiResolution Bionic Ear System M Zn-Air Battery Pak

The M Zn-Air Battery Pak has a battery door that does not require a tool to replace the button/coin cell battery.

What happened

Advanced Bionics, LLC, of Valencia, California, recalled the HiResolution Bionic Ear System M Zn-Air Battery Pak because the non-tamperproof battery door supplied with the sound processor fails to conform to IEC 60601-1-11:2015 by not requiring the use of a tool to replace the button/coin cell battery. The battery pak is a disposable power source for the Naida CI M and Sky CI M Sound Processors.

The recall began March 23, 2026, and was added to the FDA's Enforcement Report on April 27, 2026, with nationwide distribution in the U.S. and worldwide.

What to do

We don't have product-specific instructions for this recall. FDA's general advice for medical device recalls:

  1. A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
  2. Talk to your health care provider before you stop using it.

Is yours affected?

Compare the details below with your product and check the FDA notice to confirm whether it is affected.

  1. Model number
    • CI-5501-110
    • CI-5501-120
    • CI-5501-130
    • CI-5501-140
    • CI-5501-150
    • CI-5501-190
    • CI-5501-240
    • CI-5501-250
    • CI-5501-260
    • CI-5501-270
    • CI-5501-280
    Lot code
    All lots manufactured up to February 3, 2026. Refer to Exhibit_0006

Recall details

Company began the recall
Sold in

Worldwide - U.S Nationwide distribution including in the state of AK, AL, AR, AZ, CA, CO, CT, CD, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Canada, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, China, Chile, Croatia, Cuba, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Guam, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Kazakhstan, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Trinidad and Tobago, United Arab Emirates, and United Kingdom.

Company
Advanced Bionics, LLCValencia, California
Risk

Moderate risk

Using it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.

FDA Class II recall · How risk is rated

Event ID
98678
Recall number
Z-1950-2026

All recalls from Advanced Bionics, LLC

Timeline

FDA's Enforcement Report is its public list of recalls; FDA adds a recall when it posts it and publishes the list weekly.