Agiliti Adapt Pump

Agiliti is recalling the Adapt Pump (model 61600200) for a software failure that prevents the Microclimate Management function from restarting.

What to do

Contact Agiliti Health to schedule a software update to version 1.3.1 or higher to resolve the timer issue.

Recalled
Listed
Updated
Recall No.
Z-1472-2026
Agency
FDA
Firm
Agiliti Health - Ellis
Recall ID
98332
Status
ONGOING

Why it was recalled

Agiliti is recalling the Adapt Pump, model 61600200, used with Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, and Adapt Convertible EX support surfaces. The pump fails to automatically re-engage the Microclimate Management function after the 10-minute Autofirm timer elapses. This recall was distributed nationwide.

Is mine affected?

Check the model, lot, or serial number on the device or its packaging. Yours is affected if it matches one below.

Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Mattress, Air Flotation, Alternating Pressure