Adapt Pump Model 61600200
← Back to all recalls
MEDICAL DEVICE RECALLMODERATE

Adapt Pump Model 61600200

Agiliti is recalling Adapt Pump Model 61600200 due to Microclimate Management function failure after Autofirm use.

What to do

Stop using the Autofirm function on affected Adapt support surfaces. Contact Agiliti Health for a software update or replacement pump.

Posted
Updated
Recall No.
Z-1472-2026
Agency
Firm
Agiliti Health - Ellis
Recall ID
98332
Status
ONGOING

Why it was recalled

Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes. This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function.

How to identify it

UPC 0084569904914

Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Mattress, Air Flotation, Alternating Pressure

Where it was sold

Distributed in US.

← Back to all recalls

Editorial summary by Recall Canary · last updated 2026-04-27 · primary source: U.S. Food and Drug Administration · severity reflects the agency's published classification