FDA recall, , High risk
Arrow International Pressure Injectable PICC and JACC Kits
Pressure Injectable Arrowg+ard Blue Advance PICC Kits and Arrow Marked Guidewires have a coating that may swell and detach.
What to do
Check yoursIf a code on your device is listed below:
- Check the lot against the recall notice and ask your provider or supplier before you stop using it.
Typical advice for a recalled medical device; the FDA notice gives no product-specific instructions.
What happened
Arrow International, LLC of Morrisville, North Carolina is recalling several Pressure Injectable Arrowg+ard Blue Advance catheter kits, catheters and Arrow Marked Guidewires. The company started the recall on its own. The agency says the hydrophilic coated spring wire guide components may swell inconsistently when hydrated with saline.
This can cause the coating to clump, with potential for it to detach. A detached particle may be released into the blood stream, which may result in inflammation, thrombosis, ischemia, tissue injury, stroke, or further intervention.
The affected products include the Pressure Injectable Arrowg+ard Blue Advance Two-Lumen Kit pre-loaded with Arrow VPS Precision Stylet and the Arrow Marked Guidewire Hydrophilic-Coated Stainless Steel, along with other catheter and guidewire products. Customers were told by letter. The products were distributed nationwide in the United States and in countries outside the United States, including Australia, Canada, China, Japan, and others. The FDA classified the recall on November 25, 2026.
Is your kit affected?
Compare the details below with your product and check the FDA notice to confirm whether it is affected.
A UDI-DI is the device identifier printed with the barcode on the device label.
Product details
Pressure Injectable Arrowg+ard Blue Advance Two-Lumen PICC Kit pre-loaded with Arrow VPS Precision Stylet / Pressure Injectable Arrowg+ard Blue Advance One-Lumen PICC Kit pre-loaded with Arrow VPS Precision Stylet / Pressure Injectable Arrowg+ard Blue Advance One-Lumen PICC with 130 cm Hydrophilic-Coated Nitinol Guidewire
- UDI-DI
- 10801902223477
- 10801902223484
- 10801902215786
- 10801902188516
- Recall number
- Z-0498-2027
- Product
- Pressure Injectable Arrowg+ard Blue Advance Two-Lumen PICC Kit pre-loaded with Arrow VPS Precision Stylet
- Reference number
- ASK-45552-CHS
- Lot code
- ASK-45552-CHS
- 33F26A1395
- 33F26G0411
- Product
- Pressure Injectable Arrowg+ard Blue Advance One-Lumen PICC Kit pre-loaded with Arrow VPS Precision Stylet
- Reference number
- ASK-45541-CHS
- Lot code
- ASK-45541-CHS
- 33F26A1422
- Product
- Pressure Injectable Arrowg+ard Blue Advance One-Lumen PICC with 130 cm Hydrophilic-Coated Nitinol Guidewire
- Reference number
- PR-45541-HPHNL
- Lot code
- PR-45541-HPHNL
- 33F25B0399
- 33F25G0030
- 33F26A0137
- 33F26B0738
- 33F26C0475
- 33F26D0315
- 33F26E0189
- 33F26F0344
- 33F26G0135
Arrow Marked Guidewire Hydrophilic-Coated Stainless Steel / Arrow Marked Guidewire Hydrophilic-Coated Nitinol
Product details from the notice
Arrow Marked Guidewire Hydrophilic-Coated Stainless Steel / Arrow Marked Guidewire Hydrophilic-Coated Nitinol, REF: PW-16130-HN, PW-18080-HN;
- UDI-DI
- 10801902198454
- 10801902198461
- 10801902206241
- Recall number
- Z-0499-2027
- Product
- Arrow Marked Guidewire Hydrophilic-Coated Stainless Steel
- Reference number
- PW-18080-HS
- Lot code
- 9018018283
- 9018090311
- 9018264168
- 9018307037
- 9018546306
- 9018691553
- 9018756856
- 9018889316
- 9019089438
- 9019272633
- 9019454864
- 9019454877
- 9019811035
- 9020034639
- 9020074999
- 9020257269
- 9020409879
- 9020445151
- 9020641836
- 9020656340
- 9021068293
- 9021372111
- 9021578720
- 9021637861
- 9021931119
- 9021931125
- 9009092227
- 9009513080
- 9009513092
- 9009513093
- 9009520108
- 9009663277
- 9010305820
- 9010862140
- 9014671416
- 9015047855
- 9015954586
- 9016010614
- 9016235421
- 9016246159
- 9016607289
- 9016791674
- 9017079360
- 9017299757
- 9017471996
- 9017600757
- 9017746420
- 9017842480
- 9018018298
- 9018144328
All 54
- 9018264174
- 9019476228
- 9020644751
- 9020709591
- Product
- Arrow Marked Guidewire Hydrophilic-Coated Nitinol, REF: PW-16130-HN, PW-18080-HN;
- Lot code
- 9009415807
- 9009904088
- 9009904222
- 9010049063
- 9010231250
- 9010263357
- 9010385479
- 9010597691
- 9010862144
- 9014937817
- 9014965239
- 9015165211
- 9015540289
- 9017299759
- 9017600765
- 9017704999
- 9018380779
- 9018819619
- 9020445153
- 9021346112
- 9009092217
- 9009415801
- 9009415803
- 9009415804
- 9009415806
- 9009791247
- 9009791255
- 9009841780
- 9009841782
- 9009841783
- 9009862528
- 9010023487
- 9010142293
- 9010231248
- 9010385477
- 9010597685
- 9010805397
- 9014671424
- 9014695804
- 9014843191
- 9015119844
- 9015791555
- 9016851511
- 9016971267
- 9017106139
- 9017188555
- 9017261912
- 9017299806
- 9017349872
- 9017600754
All 78
- 9017669852
- 9017842462
- 9018018283
- 9018090311
- 9018264168
- 9018307037
- 9018546306
- 9018691553
- 9018756856
- 9018889316
- 9019089438
- 9019272633
- 9019454864
- 9019454877
- 9019811035
- 9020034639
- 9020074999
- 9020257269
- 9020409879
- 9020445151
- 9020641836
- 9020656340
- 9021068293
- 9021372111
- 9021578720
- 9021637861
- 9021931119
- 9021931125
Some entries omit identifying details. Check the notice to confirm which units are affected.
Product details
Pressure Injectable Arrowg+ard Blue Advance Three-Lumen JACC with 80 cm Hydrophilic-Coated Nitinol Guidewire / Pressure Injectable Arrowg+ard Blue Advance Two-Lumen JACC with 80 cm Hydrophilic-Coated Nitinol Guidewire / Pressure Injectable Arrowg+ard Blue Advance One-Lumen JACC with 80 cm Hydrophilic-Coated Nitinol Guidewire
- UDI-DI
- 10801902188646
- 10801902188653
- 10801902188387
- Recall number
- Z-0503-2027
- Product
- Pressure Injectable Arrowg+ard Blue Advance Three-Lumen JACC with 80 cm Hydrophilic-Coated Nitinol Guidewire
- Reference number
- JRT-43563-HPHNM
- Lot code
- 33F26B0312
- Product
- Pressure Injectable Arrowg+ard Blue Advance Two-Lumen JACC with 80 cm Hydrophilic-Coated Nitinol Guidewire
- Reference number
- JRT-43552-HPHNM
- Lot code
- 33F24L0169
- 33F26B0026
- 33F26C0089
- 33F26E0341
- Product
- Pressure Injectable Arrowg+ard Blue Advance One-Lumen JACC with 80 cm Hydrophilic-Coated Nitinol Guidewire
- Reference number
- JRT-43541-HPHNM
- Lot code
- 33F26B0042
Product details
Pressure Injectable Arrowg+ard Blue Advance One-Lumen PICC with 80 cm Hydrophilic-Coated Nitinol Guidewire, REF: PR-45541-HPHNM;
Pressure Injectable Arrowg+ard Blue Advance Three-Lumen PICC with 130 cm Hydrophilic-Coated Nitinol Guidewire, REF: PR-45563-HPHNL;
Pressure Injectable Arrowg+ard Blue Advance Two-Lumen PICC with 130 cm Hydrophilic-Coated Nitinol Guidewire, REF: PR-45552-HPHNL;
Pressure Injectable Arrowg+ard Blue Advance Two-Lumen PICC with 80 cm Hydrophilic-Coated Nitinol Guidewire, REF: PR-45552-HPHNM, PR-45563-HPHNM;
Pressure Injectable One-Lumen PICC, REF: PR-35541-NL;
Pressure Injectable One-Lumen PICC with 130 cm Wire, REF: PR-34041-NL, PR-35541-HPHNL;
Pressure Injectable One-Lumen PICC with 80 cm Wire, REF: PR-34041-HPHNM, PR-34041-NM, PR-35041-HPHNM, PR-35541-HPHNM;
Pressure Injectable Three-Lumen PICC, REF: PR-34063-NM, PR-35563-NL;
Pressure Injectable Three-Lumen PICC with 130 cm Wire, REF: PR-34063-NL, PR-35563-HPHNL;
Pressure Injectable Three-Lumen PICC with 80 cm Wire, REF: PR-35563-HPHNM;
Pressure Injectable Two-Lumen PICC, REF: PR-34052-NM, PR-35552-NL;
Pressure Injectable Two-Lumen PICC with 130 cm Wire, REF: PR-34052-NL, PR-35552-HPHNL;
Pressure Injectable Two-Lumen PICC with 80 cm Wire, REF: PR-34052-HPHNM, PR-35552-HPHNM- UDI-DI
- 10801902215847
- 10801902215809
- 10801902188530
- 10801902215793
- 10801902188554
- 10801902215861
- 10801902188394
- 10801902198638
- 10801902198621
- 10801902215748
- 10801902192650
- 10801902192629
- 10801902198676
- 10801902192636
- 10801902215816
- 10801902192667
- 10801902198683
- 10801902198652
- 10801902198645
- 10801902195651
- 10801902215762
- 10801902215830
- 10801902192698
- 10801902198669
- 10801902198614
- 10801902198607
- 10801902192674
- 10801902215755
- 10801902197112
- 10801902215823
- 10801902192681
- Affected units
- PR-45541-HPHNM, 33F25B0397, 33F25D0286, 33F25J1024, 33F26B0704, 33F26E0190, PR-45563-HPHNL, 33F25B0394, 33F25C0409, 33F25D0686, 33F25E0345, 33F25H0230, 33F26B0091, 33F26C0570, 33F26D0436, 33F26E0192, 33F26G0312, 33F26H0514, PR-45552-HPHNL, 33F25C0001, 33F25C0068, 33F25C0566, 33F25D0022, 33F25K0383, 33F25M0506, 33F26C0055, 33F26D0070, 33F26D0553, 33F26E0582, 33F26G0136, 33F26G1061, 33F26H0677, 33S24B0614, 33S25B0614, PR-45552-HPHNM, 33F25B0395, 33F25B0402, 33F25C0059, 33F25D0075, 33F25H0305, 33F26A0045, 33F26C0476, 33F26C0523, 33F26D0672, 33F26E0188, 33F26F0608, 33F26G0134, 33F26G0458, 33F26G0956, 33F26H0219, PR-45563-HPHNM, 33F25B0401, 33F25C0067, 33F25D0002, 33F25D0101, 33F25G0031, 33F25H0831, 33F26A1274, 33F26C0506, 33F26C0526, PR-35541-NL, 33F25B0106, 33F25C0408, 33F25D0163, 33F25E0346, 33F25E0393, 33F25H0344, 33F25H0692, 33F25H0863, 33F25J0072, PR-34041-NL, 33F25A0359, 33F25B0303, 33F25C0065, 33F25F0001, 33F25J0006, PR-35541-HPHNL, 33F25A0049, 33F25B0103, 33F25D0162, 33F25E0324, 33F25E0355, 33F25F0340, 33F25H0253, 33F25H0675, 33F25J0007, 33F25J0074, 33F25J0972, 33F25K0012, 33F25M0835, 33F26A1271, 33F26D0294, 33F26D0301, 33F26E0191, 33F26F0109, 33F26G0618, 33S25D0471, PR-34041-HPHNM, 33F25B0217, 33F25C0082, 33F25F0016, 33F25F0261, 33F25G0626, 33F25H0301, 33F25H0681, 33F25H0792, 33F25K0241, 33F26A1276, 33F26C0471, 33S25G0800, 33S25H0467, 33S25H0997, PR-34041-NM, 33F25B0286, 33F25E0459, 33F25F0327, PR-35041-HPHNM, 33F25A0445, 33F25B0007, 33F25B0049, 33F25B0324, 33F25D0105, 33F25D0177, 33F25E0353, 33F25E1106, 33F25H0228, 33F25H0862, 33F25J0994, 33F25K0223, 33F25M0780, 33F26A1263, 33F26C0092, 33F26E0143, PR-35541-HPHNM, 33F24M0142, 33F25B0050, 33F25C0077, 33F25C0385, 33F25D0132, 33F25D0179, 33F25E1135, 33F25H0303, 33F26A1317, 33F26B0700, 33F26E0012, 33F26G0611, 33F26G0952, PR-34063-NM, 33F25C0423, 33F25D0161, 33F25E0426, PR-35563-NL, 33F25B0012, 33F25B0107, 33F25B0352, 33F25D0025, 33F25D0183, 33F25H0256, 33F25J0071, PR-34063-NL, 33F25B0071, 33F25C0221, PR-35563-HPHNL, 33F25A0091, 33F25B0122, 33F25B0411, 33F25C0116, 33F25C0599, 33F25F0342, 33F25G0726, 33F25H0673, 33F25J0002, 33F25K0014, 33F26F0134, PR-35563-HPHNM, 33F24H0294, 33F25B0022, 33F25E0460, 33F25F0387, 33F25H0830, 33F25J0973, 33F25K0270, 33F25L0009, 33F25M0829, 33F26C0472, 33F26D0653, 33F26F0140, PR-34052-NM, 33F25B0101, 33F25B0224, 33F25B0349, 33F25C0537, 33F25E0319, 33F25G0619, PR-35552-NL, 33F25B0121, 33F25B0222, 33F25C0485, 33F25D0079, 33F25F0013, 33F25F0307, 33F25H0220, 33F25H0304, 33F25J0989, 33F25K0287, 33F25M0041, 33F26A0042, PR-34052-NL, 33F25A0379, 33F25B0070, 33F25C0571, 33F25D0160, 33F25E0318, PR-35552-HPHNL, 33F24L0030, 33F25A0065, 33F25B0030, 33F25B0325, 33F25B0351, 33F25D0095, 33F25D0180, 33F25F0341, 33F25H0255, 33F25K0013, 33F25M0826, 33F26A1269, 33F26E0216, 33F26E0583, 33F26F0108, 33F26G0138, 33F26G0950, PR-34052-HPHNM, 33F25A0057, 33F25B0100, 33F25B0348, 33F25C0514, 33F25D0176, 33F25F0305, 33F25G0426, 33F25G0606, 33F25H0861, 33F25J0968, 33F25L0291, 33F26B0310, 33F26C0091, 33F26E0581, 33F26F0564, PR-35552-HPHNM, 33F24L0529, 33F25C0403, 33F25C0573, 33F25D0016, 33F25E0368, 33F25J0956, 33F25L0048, 33F25L0049, 33F26A0363, 33F26B0740, 33F26C0505, 33F26D0295, 33F26D0463, 33F26G0958, 33F26H0057, 33F26H0221, 33S25F0436
Pressure Injectable Arrowg+ard Blue Advance Two-Lumen PICC Kit pre-loaded with Arrow VPS Precision Stylet
- UDI-DI
- 10801902223347
- Reference number
- ASK-45052-MC
- Lot code
- 33F25G0463
- 33F26D0568
Pressure Injectable Three-Lumen PICC with 80 cm Wire
Product details from the notice
Pressure Injectable Three-Lumen PICC with 80 cm Wire, REF: PR-34063-Hphnm, PR-35052-Hphnm
- UDI-DI
- 10801902195637
- 10801902192643
- Lot code
- 33F25A0361
- 33F25B0323
- 33F25E1160
- 33F26B0046
- 33F24M0060
- 33F24M0083
- 33F25A0372
- 33F25B0072
- 33F25C0516
- 33F25C0560
- 33F25C0602
- 33F25D0024
- 33F25D0074
- 33F25D0212
- 33F25E0323
- 33F25E0336
- 33F25E0351
- 33F25F0470
- 33F25G0644
- 33F25H0343
- 33F25H0716
- 33F25H0874
- 33F25J1046
- 33F25K0217
- 33F25M0505
- 33F26A0259
- 33F26A1286
- 33F26B0691
- 33F26C0715
- 33F26D0270
- 33F26D0427
- 33F26E0621
- 33F26E0708
Recall details
- Posted by FDA
- Recall began
- Status
- Open
- Weekly archive
- Listed the week of Nov 23, 2026
- Sold in
(US nationwide) California, South Carolina, Georgia, Arizona, Colorado, Missouri, Louisiana, Arkansas, Texas, Utah, Montana, Iowa, Massachusetts, Alabama, Wyoming, New Jersey, Virginia, Pennsylvania, West Virginia, Delaware, Ohio, Nebraska, Indiana, Florida, Tennessee, New York, New Hampshire, Mississippi, Maryland, Michigan, Kansas, New Mexico, Kentucky, Oklahoma, Maine, North Carolina, Idaho, Washington, Hawaii, Nevada, Rhode Island, Oregon, Vermont, Illinois, Wisconsin, South Dakota, Minnesota, District of Columbia, Connecticut; (outside the U.S. countries of): Australia, Brunei, Canada, Chile, China, Colombia, India, Japan, South Korea, Mexico, Malaysia, Nepal, New Zealand, Philippines, Singapore, Thailand, Taiwan, and Vietnam.
- Recalled by
- Arrow International, LLCMorrisville, North Carolina
- Hazard
- Foreign object
- Risk
High risk
There is a reasonable chance that using it could cause serious health problems or death.
FDA Class I recall · How risk is rated
- FDA event number
- 100262
- Recall numbers
- Z-0498-2027 to Z-0503-2027
- Recall type
- Started by the company
- How the company first told customers
- Letter
All recalls from Arrow International, LLC
Timeline
FDA's Enforcement Report is its public list of recalls; FDA posts a recall by adding it there.
ARROW INTERNATIONAL began the recall
FDA posted the recall to its Enforcement Report and rated it Class I (high risk)
Added to Recall Canary