FDA recall, , High risk

Arrow International Pressure Injectable PICC and JACC Kits

Pressure Injectable Arrowg+ard Blue Advance PICC Kits and Arrow Marked Guidewires have a coating that may swell and detach.

Product photo

What to do

Check yours

If a code on your device is listed below:

  1. Check the lot against the recall notice and ask your provider or supplier before you stop using it.

Typical advice for a recalled medical device; the FDA notice gives no product-specific instructions.

What happened

Arrow International, LLC of Morrisville, North Carolina is recalling several Pressure Injectable Arrowg+ard Blue Advance catheter kits, catheters and Arrow Marked Guidewires. The company started the recall on its own. The agency says the hydrophilic coated spring wire guide components may swell inconsistently when hydrated with saline.

This can cause the coating to clump, with potential for it to detach. A detached particle may be released into the blood stream, which may result in inflammation, thrombosis, ischemia, tissue injury, stroke, or further intervention.

The affected products include the Pressure Injectable Arrowg+ard Blue Advance Two-Lumen Kit pre-loaded with Arrow VPS Precision Stylet and the Arrow Marked Guidewire Hydrophilic-Coated Stainless Steel, along with other catheter and guidewire products. Customers were told by letter. The products were distributed nationwide in the United States and in countries outside the United States, including Australia, Canada, China, Japan, and others. The FDA classified the recall on November 25, 2026.

Is your kit affected?

Compare the details below with your product and check the FDA notice to confirm whether it is affected.

A UDI-DI is the device identifier printed with the barcode on the device label.

  1. Product details

    Pressure Injectable Arrowg+ard Blue Advance Two-Lumen PICC Kit pre-loaded with Arrow VPS Precision Stylet / Pressure Injectable Arrowg+ard Blue Advance One-Lumen PICC Kit pre-loaded with Arrow VPS Precision Stylet / Pressure Injectable Arrowg+ard Blue Advance One-Lumen PICC with 130 cm Hydrophilic-Coated Nitinol Guidewire

    UDI-DI
    • 10801902223477
    • 10801902223484
    • 10801902215786
    • 10801902188516
    Recall number
    Z-0498-2027
    1. Product
      Pressure Injectable Arrowg+ard Blue Advance Two-Lumen PICC Kit pre-loaded with Arrow VPS Precision Stylet
      Reference number
      • ASK-45552-CHS
      Lot code
      • ASK-45552-CHS
      • 33F26A1395
      • 33F26G0411
    2. Product
      Pressure Injectable Arrowg+ard Blue Advance One-Lumen PICC Kit pre-loaded with Arrow VPS Precision Stylet
      Reference number
      • ASK-45541-CHS
      Lot code
      • ASK-45541-CHS
      • 33F26A1422
    3. Product
      Pressure Injectable Arrowg+ard Blue Advance One-Lumen PICC with 130 cm Hydrophilic-Coated Nitinol Guidewire
      Reference number
      • PR-45541-HPHNL
      Lot code
      • PR-45541-HPHNL
      • 33F25B0399
      • 33F25G0030
      • 33F26A0137
      • 33F26B0738
      • 33F26C0475
      • 33F26D0315
      • 33F26E0189
      • 33F26F0344
      • 33F26G0135
  2. Arrow Marked Guidewire Hydrophilic-Coated Stainless Steel / Arrow Marked Guidewire Hydrophilic-Coated Nitinol

    Product details from the notice

    Arrow Marked Guidewire Hydrophilic-Coated Stainless Steel / Arrow Marked Guidewire Hydrophilic-Coated Nitinol, REF: PW-16130-HN, PW-18080-HN;

    UDI-DI
    • 10801902198454
    • 10801902198461
    • 10801902206241
    Recall number
    Z-0499-2027
    1. Product
      Arrow Marked Guidewire Hydrophilic-Coated Stainless Steel
      Reference number
      • PW-18080-HS
      Lot code
      • 9018018283
      • 9018090311
      • 9018264168
      • 9018307037
      • 9018546306
      • 9018691553
      • 9018756856
      • 9018889316
      • 9019089438
      • 9019272633
      • 9019454864
      • 9019454877
      • 9019811035
      • 9020034639
      • 9020074999
      • 9020257269
      • 9020409879
      • 9020445151
      • 9020641836
      • 9020656340
      • 9021068293
      • 9021372111
      • 9021578720
      • 9021637861
      • 9021931119
      • 9021931125
      • 9009092227
      • 9009513080
      • 9009513092
      • 9009513093
      • 9009520108
      • 9009663277
      • 9010305820
      • 9010862140
      • 9014671416
      • 9015047855
      • 9015954586
      • 9016010614
      • 9016235421
      • 9016246159
      • 9016607289
      • 9016791674
      • 9017079360
      • 9017299757
      • 9017471996
      • 9017600757
      • 9017746420
      • 9017842480
      • 9018018298
      • 9018144328
      All 54
      • 9018264174
      • 9019476228
      • 9020644751
      • 9020709591
    2. Product
      Arrow Marked Guidewire Hydrophilic-Coated Nitinol, REF: PW-16130-HN, PW-18080-HN;
      Lot code
      • 9009415807
      • 9009904088
      • 9009904222
      • 9010049063
      • 9010231250
      • 9010263357
      • 9010385479
      • 9010597691
      • 9010862144
      • 9014937817
      • 9014965239
      • 9015165211
      • 9015540289
      • 9017299759
      • 9017600765
      • 9017704999
      • 9018380779
      • 9018819619
      • 9020445153
      • 9021346112
      • 9009092217
      • 9009415801
      • 9009415803
      • 9009415804
      • 9009415806
      • 9009791247
      • 9009791255
      • 9009841780
      • 9009841782
      • 9009841783
      • 9009862528
      • 9010023487
      • 9010142293
      • 9010231248
      • 9010385477
      • 9010597685
      • 9010805397
      • 9014671424
      • 9014695804
      • 9014843191
      • 9015119844
      • 9015791555
      • 9016851511
      • 9016971267
      • 9017106139
      • 9017188555
      • 9017261912
      • 9017299806
      • 9017349872
      • 9017600754
      All 78
      • 9017669852
      • 9017842462
      • 9018018283
      • 9018090311
      • 9018264168
      • 9018307037
      • 9018546306
      • 9018691553
      • 9018756856
      • 9018889316
      • 9019089438
      • 9019272633
      • 9019454864
      • 9019454877
      • 9019811035
      • 9020034639
      • 9020074999
      • 9020257269
      • 9020409879
      • 9020445151
      • 9020641836
      • 9020656340
      • 9021068293
      • 9021372111
      • 9021578720
      • 9021637861
      • 9021931119
      • 9021931125

    Some entries omit identifying details. Check the notice to confirm which units are affected.

  3. Product details

    Pressure Injectable Arrowg+ard Blue Advance Three-Lumen JACC with 80 cm Hydrophilic-Coated Nitinol Guidewire / Pressure Injectable Arrowg+ard Blue Advance Two-Lumen JACC with 80 cm Hydrophilic-Coated Nitinol Guidewire / Pressure Injectable Arrowg+ard Blue Advance One-Lumen JACC with 80 cm Hydrophilic-Coated Nitinol Guidewire

    UDI-DI
    • 10801902188646
    • 10801902188653
    • 10801902188387
    Recall number
    Z-0503-2027
    1. Product
      Pressure Injectable Arrowg+ard Blue Advance Three-Lumen JACC with 80 cm Hydrophilic-Coated Nitinol Guidewire
      Reference number
      • JRT-43563-HPHNM
      Lot code
      • 33F26B0312
    2. Product
      Pressure Injectable Arrowg+ard Blue Advance Two-Lumen JACC with 80 cm Hydrophilic-Coated Nitinol Guidewire
      Reference number
      • JRT-43552-HPHNM
      Lot code
      • 33F24L0169
      • 33F26B0026
      • 33F26C0089
      • 33F26E0341
    3. Product
      Pressure Injectable Arrowg+ard Blue Advance One-Lumen JACC with 80 cm Hydrophilic-Coated Nitinol Guidewire
      Reference number
      • JRT-43541-HPHNM
      Lot code
      • 33F26B0042
  4. Product details

    Pressure Injectable Arrowg+ard Blue Advance One-Lumen PICC with 80 cm Hydrophilic-Coated Nitinol Guidewire, REF: PR-45541-HPHNM;
    Pressure Injectable Arrowg+ard Blue Advance Three-Lumen PICC with 130 cm Hydrophilic-Coated Nitinol Guidewire, REF: PR-45563-HPHNL;
    Pressure Injectable Arrowg+ard Blue Advance Two-Lumen PICC with 130 cm Hydrophilic-Coated Nitinol Guidewire, REF: PR-45552-HPHNL;
    Pressure Injectable Arrowg+ard Blue Advance Two-Lumen PICC with 80 cm Hydrophilic-Coated Nitinol Guidewire, REF: PR-45552-HPHNM, PR-45563-HPHNM;
    Pressure Injectable One-Lumen PICC, REF: PR-35541-NL;
    Pressure Injectable One-Lumen PICC with 130 cm Wire, REF: PR-34041-NL, PR-35541-HPHNL;
    Pressure Injectable One-Lumen PICC with 80 cm Wire, REF: PR-34041-HPHNM, PR-34041-NM, PR-35041-HPHNM, PR-35541-HPHNM;
    Pressure Injectable Three-Lumen PICC, REF: PR-34063-NM, PR-35563-NL;
    Pressure Injectable Three-Lumen PICC with 130 cm Wire, REF: PR-34063-NL, PR-35563-HPHNL;
    Pressure Injectable Three-Lumen PICC with 80 cm Wire, REF: PR-35563-HPHNM;
    Pressure Injectable Two-Lumen PICC, REF: PR-34052-NM, PR-35552-NL;
    Pressure Injectable Two-Lumen PICC with 130 cm Wire, REF: PR-34052-NL, PR-35552-HPHNL;
    Pressure Injectable Two-Lumen PICC with 80 cm Wire, REF: PR-34052-HPHNM, PR-35552-HPHNM

    UDI-DI
    • 10801902215847
    • 10801902215809
    • 10801902188530
    • 10801902215793
    • 10801902188554
    • 10801902215861
    • 10801902188394
    • 10801902198638
    • 10801902198621
    • 10801902215748
    • 10801902192650
    • 10801902192629
    • 10801902198676
    • 10801902192636
    • 10801902215816
    • 10801902192667
    • 10801902198683
    • 10801902198652
    • 10801902198645
    • 10801902195651
    • 10801902215762
    • 10801902215830
    • 10801902192698
    • 10801902198669
    • 10801902198614
    • 10801902198607
    • 10801902192674
    • 10801902215755
    • 10801902197112
    • 10801902215823
    • 10801902192681
    Affected units
    PR-45541-HPHNM, 33F25B0397, 33F25D0286, 33F25J1024, 33F26B0704, 33F26E0190, PR-45563-HPHNL, 33F25B0394, 33F25C0409, 33F25D0686, 33F25E0345, 33F25H0230, 33F26B0091, 33F26C0570, 33F26D0436, 33F26E0192, 33F26G0312, 33F26H0514, PR-45552-HPHNL, 33F25C0001, 33F25C0068, 33F25C0566, 33F25D0022, 33F25K0383, 33F25M0506, 33F26C0055, 33F26D0070, 33F26D0553, 33F26E0582, 33F26G0136, 33F26G1061, 33F26H0677, 33S24B0614, 33S25B0614, PR-45552-HPHNM, 33F25B0395, 33F25B0402, 33F25C0059, 33F25D0075, 33F25H0305, 33F26A0045, 33F26C0476, 33F26C0523, 33F26D0672, 33F26E0188, 33F26F0608, 33F26G0134, 33F26G0458, 33F26G0956, 33F26H0219, PR-45563-HPHNM, 33F25B0401, 33F25C0067, 33F25D0002, 33F25D0101, 33F25G0031, 33F25H0831, 33F26A1274, 33F26C0506, 33F26C0526, PR-35541-NL, 33F25B0106, 33F25C0408, 33F25D0163, 33F25E0346, 33F25E0393, 33F25H0344, 33F25H0692, 33F25H0863, 33F25J0072, PR-34041-NL, 33F25A0359, 33F25B0303, 33F25C0065, 33F25F0001, 33F25J0006, PR-35541-HPHNL, 33F25A0049, 33F25B0103, 33F25D0162, 33F25E0324, 33F25E0355, 33F25F0340, 33F25H0253, 33F25H0675, 33F25J0007, 33F25J0074, 33F25J0972, 33F25K0012, 33F25M0835, 33F26A1271, 33F26D0294, 33F26D0301, 33F26E0191, 33F26F0109, 33F26G0618, 33S25D0471, PR-34041-HPHNM, 33F25B0217, 33F25C0082, 33F25F0016, 33F25F0261, 33F25G0626, 33F25H0301, 33F25H0681, 33F25H0792, 33F25K0241, 33F26A1276, 33F26C0471, 33S25G0800, 33S25H0467, 33S25H0997, PR-34041-NM, 33F25B0286, 33F25E0459, 33F25F0327, PR-35041-HPHNM, 33F25A0445, 33F25B0007, 33F25B0049, 33F25B0324, 33F25D0105, 33F25D0177, 33F25E0353, 33F25E1106, 33F25H0228, 33F25H0862, 33F25J0994, 33F25K0223, 33F25M0780, 33F26A1263, 33F26C0092, 33F26E0143, PR-35541-HPHNM, 33F24M0142, 33F25B0050, 33F25C0077, 33F25C0385, 33F25D0132, 33F25D0179, 33F25E1135, 33F25H0303, 33F26A1317, 33F26B0700, 33F26E0012, 33F26G0611, 33F26G0952, PR-34063-NM, 33F25C0423, 33F25D0161, 33F25E0426, PR-35563-NL, 33F25B0012, 33F25B0107, 33F25B0352, 33F25D0025, 33F25D0183, 33F25H0256, 33F25J0071, PR-34063-NL, 33F25B0071, 33F25C0221, PR-35563-HPHNL, 33F25A0091, 33F25B0122, 33F25B0411, 33F25C0116, 33F25C0599, 33F25F0342, 33F25G0726, 33F25H0673, 33F25J0002, 33F25K0014, 33F26F0134, PR-35563-HPHNM, 33F24H0294, 33F25B0022, 33F25E0460, 33F25F0387, 33F25H0830, 33F25J0973, 33F25K0270, 33F25L0009, 33F25M0829, 33F26C0472, 33F26D0653, 33F26F0140, PR-34052-NM, 33F25B0101, 33F25B0224, 33F25B0349, 33F25C0537, 33F25E0319, 33F25G0619, PR-35552-NL, 33F25B0121, 33F25B0222, 33F25C0485, 33F25D0079, 33F25F0013, 33F25F0307, 33F25H0220, 33F25H0304, 33F25J0989, 33F25K0287, 33F25M0041, 33F26A0042, PR-34052-NL, 33F25A0379, 33F25B0070, 33F25C0571, 33F25D0160, 33F25E0318, PR-35552-HPHNL, 33F24L0030, 33F25A0065, 33F25B0030, 33F25B0325, 33F25B0351, 33F25D0095, 33F25D0180, 33F25F0341, 33F25H0255, 33F25K0013, 33F25M0826, 33F26A1269, 33F26E0216, 33F26E0583, 33F26F0108, 33F26G0138, 33F26G0950, PR-34052-HPHNM, 33F25A0057, 33F25B0100, 33F25B0348, 33F25C0514, 33F25D0176, 33F25F0305, 33F25G0426, 33F25G0606, 33F25H0861, 33F25J0968, 33F25L0291, 33F26B0310, 33F26C0091, 33F26E0581, 33F26F0564, PR-35552-HPHNM, 33F24L0529, 33F25C0403, 33F25C0573, 33F25D0016, 33F25E0368, 33F25J0956, 33F25L0048, 33F25L0049, 33F26A0363, 33F26B0740, 33F26C0505, 33F26D0295, 33F26D0463, 33F26G0958, 33F26H0057, 33F26H0221, 33S25F0436
  5. Pressure Injectable Arrowg+ard Blue Advance Two-Lumen PICC Kit pre-loaded with Arrow VPS Precision Stylet

    UDI-DI
    • 10801902223347
    Reference number
    • ASK-45052-MC
    Lot code
    • 33F25G0463
    • 33F26D0568
  6. Pressure Injectable Three-Lumen PICC with 80 cm Wire

    Product details from the notice

    Pressure Injectable Three-Lumen PICC with 80 cm Wire, REF: PR-34063-Hphnm, PR-35052-Hphnm

    UDI-DI
    • 10801902195637
    • 10801902192643
    Lot code
    • 33F25A0361
    • 33F25B0323
    • 33F25E1160
    • 33F26B0046
    • 33F24M0060
    • 33F24M0083
    • 33F25A0372
    • 33F25B0072
    • 33F25C0516
    • 33F25C0560
    • 33F25C0602
    • 33F25D0024
    • 33F25D0074
    • 33F25D0212
    • 33F25E0323
    • 33F25E0336
    • 33F25E0351
    • 33F25F0470
    • 33F25G0644
    • 33F25H0343
    • 33F25H0716
    • 33F25H0874
    • 33F25J1046
    • 33F25K0217
    • 33F25M0505
    • 33F26A0259
    • 33F26A1286
    • 33F26B0691
    • 33F26C0715
    • 33F26D0270
    • 33F26D0427
    • 33F26E0621
    • 33F26E0708

Recall details

Posted by FDA
Recall began
Status
Open
Sold in

(US nationwide) California, South Carolina, Georgia, Arizona, Colorado, Missouri, Louisiana, Arkansas, Texas, Utah, Montana, Iowa, Massachusetts, Alabama, Wyoming, New Jersey, Virginia, Pennsylvania, West Virginia, Delaware, Ohio, Nebraska, Indiana, Florida, Tennessee, New York, New Hampshire, Mississippi, Maryland, Michigan, Kansas, New Mexico, Kentucky, Oklahoma, Maine, North Carolina, Idaho, Washington, Hawaii, Nevada, Rhode Island, Oregon, Vermont, Illinois, Wisconsin, South Dakota, Minnesota, District of Columbia, Connecticut; (outside the U.S. countries of): Australia, Brunei, Canada, Chile, China, Colombia, India, Japan, South Korea, Mexico, Malaysia, Nepal, New Zealand, Philippines, Singapore, Thailand, Taiwan, and Vietnam.

Recalled by
Arrow International, LLCMorrisville, North Carolina
Risk

High risk

There is a reasonable chance that using it could cause serious health problems or death.

FDA Class I recall · How risk is rated

FDA event number
100262
Recall numbers
Z-0498-2027 to Z-0503-2027
Recall type
Started by the company
How the company first told customers
Letter

All recalls from Arrow International, LLC

Timeline

FDA's Enforcement Report is its public list of recalls; FDA posts a recall by adding it there.

  1. ARROW INTERNATIONAL began the recall

  2. FDA posted the recall to its Enforcement Report and rated it Class I (high risk)

  3. Added to Recall Canary

Summary by Recall Canary, written with AI · summary updated Nov 26, 2026 · Official source: U.S. Food and Drug Administration (opens in a new tab) recall · cite: “Arrow International Pressure Injectable PICC and JACC Kits Recall.” Recall Canary. https://recallcanary.com/recall/fda/arrow-pressure-injectable-arrowg-ard-blue-advance-one-lumen-picc-kit.