Welch Allyn Spot Vision Screener VS100

Welch Allyn is recalling the Spot Vision Screener VS100 (software v3.2.0.1) due to a screen flickering issue that may trigger seizures.

What to do

Contact your local Welch Allyn representative or visit the company's website to obtain the necessary software update to resolve the flickering issue.

Recalled
Listed
Updated
Recall No.
Z-1573-2026
Agency
FDA
Firm
Baxter Healthcare Corporation
Recall ID
98439
Status
ONGOING

Why it was recalled

Welch Allyn is recalling the Spot Vision Screener VS100, software version 3.2.0.1, because the display screen may unintentionally flicker. This flickering poses a risk of inducing photosensitive seizure activity in susceptible individuals. The product was distributed nationwide.

Is mine affected?

Check the model, lot, or serial number on the device or its packaging. Yours is affected if it matches one below.

Welch Allyn, Inc. Welch Allyn Spot Vision Screener VS100

UDI-DI
00732094214154, 00732094212228
Software version
Software version 3.2.0.1

Where it was sold

Distributed nationwide.