MEDICAL DEVICE RECALLMODERATE

Boston Scientific Rapid Refill Continuous Injection System

Boston Scientific is recalling the Rapid Refill Continuous Injection System due to a risk of syringe disconnection during use.

What to do

Healthcare providers should immediately stop using the affected devices and contact Boston Scientific for return and replacement instructions.

Posted
Updated
Recall No.
Z-2725-2026
Agency
Firm
Boston Scientific Corporation
Recall ID
99336
Status
ONGOING

Why it was recalled

Boston Scientific is recalling the Rapid Refill Continuous Injection System because it contains a Medline Namic Angiographic Control Syringe with a rotating adaptor that may loosen or disconnect from the manifold. This defect can cause fluid leakage, difficulty with aspiration or injection, or complete separation of the syringe from the system, which may require a device exchange and prolong medical procedures.

Is mine affected?

Check the model, lot, or serial number on the device or its packaging. Yours is affected if it matches one below.

Rapid Refill Continuous Injection System, UPN M00566001

UPC
08714729285151, 33786611, 33825949, 33826026, 33949916, 33968571, 33968575, 34079981, 34153104, 34153413, 34153415, 34203075, 34203077, 34153411, 34203079, 34203115, 34386605, 34386609, 34386613, 34429656
Recall #
Z-2725-2026

Where it was sold

Distributed nationwide.