Skip to main content
Recall Canary
TheraSphere 360 Y-90 Management Platform
← Back to all recalls
MEDICAL DEVICE RECALLMODERATE

TheraSphere 360 Y-90 Management Platform

Boston Scientific recalls TheraSphere 360 Y-90 Platform due to software anomaly causing incorrect dose calculations.

What to do

Contact Boston Scientific for software update instructions. Do not use the system until the update is applied.

Posted
Updated
Agency
Firm
Boston Scientific Corporation
Recall ID
99236
Status
ONGOING

Why it was recalled

Boston Scientific is recalling the TheraSphere 360 Y-90 Management Platform, a radiological image processing system, due to a software anomaly. When a user modifies input in the Activity Calculator and recalculates, the dose calculation updates correctly but previously added vials are not automatically updated to reflect the revised calculation, potentially leading to incorrect dose administration.

Is mine affected?

This recall covers the product below. If you have it, yours is affected.

Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Processing System.

Other recent recalls from Boston Scientific Corporation

← Back to all recalls

Editorial summary by Recall Canary · last updated 2026-07-23 · primary source: U.S. Food and Drug Administration · severity reflects the agency's published classification