Skip to main content

Ongoing FDAEvent 99817 (8 days ago)

Canon Medical Alphenix INFX-8000C/B5

The Alphenix INFX-8000C/B5 has bolts that may loosen, causing the tabletop to shift and misalignment.

Moderate risk

Using it may cause temporary or medically reversible health problems. Serious harm is unlikely.

FDA Class II recall · How risk is rated

What to do

What happened

Canon Medical System, USA, INC. recalled the Alphenix INFX-8000C/B5 after the FDA noted that loosening of the bolts securing the longitudinal movement mechanism may cause the tabletop to shift, resulting in positional, imaging, and patient misalignment and a longitudinal motion error. The recall of one unit began on August 26, 2026, and was added to the FDA's Enforcement Report on October 7, 2026.

Summary by Recall Canary from the FDA notice.

What to do

We don't have product-specific instructions for this recall. FDA's general advice for medical device recalls:

  1. A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
  2. Talk to your health care provider before you stop using it.

Is yours affected?

Affected if its UDI-DI matches.

  1. Alphenix INFX-8000C/B5

    UDI-DI
    • 4987670100192

Details

Company began the recall
Event ID
99817
Company
Canon Medical System, USA, INC.
Quantity recalled
1 unit
  1. Company began the recall

  2. Added to FDA's Enforcement Report · Classified by FDA

  3. Listed on Recall Canary

  4. In FDA's weekly enforcement report

All recalls from Canon Medical System, USA, INC.

Summary by Recall Canary, written with AI · summary updated Oct 8, 2026 · primary source: U.S. Food and Drug Administration