Skip to main content

FDA recall, , Moderate risk

Canon Medical Vantage Galan 3T and Vantage Titan 3T MRI Systems

If a quench occurs, helium gas may not escape, and pressure in the helium vessel may rise.

What happened

Canon Medical System, USA, INC., of Tustin, California, is recalling certain Vantage Titan 3T and Vantage Galan 3T MRI systems because of a potential for ice formation in the venting system of the superconducting magnet.

If a quench occurs in this condition, helium gas may not escape via the designed vent paths, and pressure in the helium vessel may rise, possibly rupturing the vessel and releasing helium gas into the MR scanning room. The recall began on December 19, 2025, and was added to the FDA's Enforcement Report on January 13, 2026.

The affected systems are the Vantage Titan 3T Model MRT-3010/MEXL-3010 and the Vantage Galan 3T Model MRT-3020/MEXL-3020, distributed nationwide in states including California, Colorado, Florida, Iowa, Illinois, Indiana, Kansas, Maryland, Missouri, Montana, Nevada, New York, Ohio, Oregon, Pennsylvania, Texas, Virginia, and Wisconsin.

What to do

We don't have product-specific instructions for this recall. FDA's general advice for medical device recalls:

  1. A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
  2. Talk to your health care provider before you stop using it.

Is yours affected?

2 products

Affected if its UDI-DI is listed.

A UDI-DI is the device identifier printed with the barcode on the device label.

  1. Vantage Titan 3T Model MRT-3010/MEXL-3010

    UDI-DI
    • 04987670101649
    • 30000041
    • 30001708
    • 30005894
    • 30007620
    • 30019175
    • 30024266
    • 30027318
    • 30029482
    • 30050314
  2. Vantage Galan 3T Model MRT-3020/MEXL-3020

    UDI-DI
    • 04987670102615
    • 30058153
    • 30075456
    • 30097122
    • 30047848
    • 30102813
    • 30108012
    • 30100563
    • 30060527
    • 30064817
    • 30069306
    • 30071480
    • 30072643
    • 30085151
    • 30098384
    • 30100803
    • 30109383
    • 30007763
    • 30031740
    • 30034832
    • 30036503
    • 30040917
    • 30041962
    • 30042870
    • 30043510
    • 30053897
    • 30056987
    • 30087647
    • 30087649
    • 30092401
    • 30056208
    • 30044596
    • 30081608
    • 30094084
    • 30119571

Recall details

Company began the recall
Sold in

U.S. Nationwide distribution in the states of California, Colorado, Florida, Iowa, Illinois, Indiana, Kansas, Maryland, Missouri, Montana, Nevada, New York, Ohio, Oregon, Pennsylvania, PR, Texas, Virginia, and Wisconsin.

Company
Canon Medical System, USA, INC.Tustin, California
Risk

Moderate risk

Using it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.

FDA Class II recall · How risk is rated

Event ID
98112
Recall numbers
Z-1017-2026 to Z-1018-2026

All recalls from Canon Medical System, USA, INC.

Timeline

FDA's Enforcement Report is its public list of recalls; FDA adds a recall when it posts it and publishes the list weekly.

  1. Company began the recall

  2. Added to FDA's Enforcement Report · Classified by FDA

  3. In FDA's weekly enforcement report

  4. Listed on Recall Canary