MEDICAL DEVICE RECALLMODERATE

WASPLab WebApp Software

Copan WASP recalls WASPLab WebApp software over a result-overwrite issue.

What to do

Contact Copan WASP to obtain the corrected software version and instructions for updating your WASPLab device.

Recalled
Listed
Updated
Recall No.
Z-2854-2026
Agency
FDA
Firm
Copan WASP
Recall ID
99302
Status
ONGOING

Why it was recalled

Copan WASP is recalling WASPLab WebApp software versions 2.33.1, 2.33.3, and 2.34.2 due to a software issue that may prevent users from overwriting results, potentially leading to incorrect microbiological test outcomes.

Is mine affected?

Check the model, lot, or serial number on the device or its packaging. Yours is affected if it matches one below.

WASPLab

Lot code
2.33.1, 2.33.3, 2.34.2
Model/Catalog number
W087
Software version
WebApp version 2.33.1, 2.33.3 and 2.34.2

Where it was sold

Distributed nationwide.

Timeline