Skip to main content

FDA recall, , Moderate risk

Corin Unity Knee ASC Modular Handle

The Unity Knee ASC Modular Handle may malfunction, causing its locking mechanism to fail.

What happened

CORIN MEDICAL, LTD. of Cirencester, United Kingdom, is voluntarily recalling the Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. The recall was initiated on October 2, 2026, and added to the FDA's Enforcement Report on November 3, 2026.

The potential malfunction of the Modular Handles is where the locking mechanism may fail, and any mating instruments already engaged to the handle will not be able to be disengaged, and the handle will be unable to fitted securely to any unengaged mating instruments.

The recall affects 22 units distributed worldwide, including in the U.S. states of California, Georgia, Illinois, Michigan, North Carolina, Rhode Island, Texas, and Utah, and the countries of Belgium, France, Germany, and Hungary.

Customers were notified by letter. The recall was classified by the FDA on November 3, 2026.

What to do

The FDA notice gives no instructions for this recall. FDA's general advice for medical device recalls:

  1. A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
  2. Talk to your health care provider before you stop using it.

Is yours affected?

Match the lot code and part code listed under your product.

  1. Unity Knee ASC Modular Handle included in the Corin Unity ASC Instrument System

    Part code
    • 1232040.
    Lot code
    • 097921

Recall details

Posted by FDA
Recall began
Status
Open
Sold in

Worldwide - US Nationwide distribution in the states of California, Georgia, Illinois, Michigan, North Carolina, Rhode Island, Texas, Utah and the countries of Belgium, France, Germany, Hungary.

Recalled by
CORIN MEDICAL, LTD.Cirencester, United Kingdom
Risk

Moderate risk

Using it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.

FDA Class II recall · How risk is rated

FDA event number
99959
Recall number
Z-0356-2027
Quantity recalled
22 units
Recall type
Voluntary: Firm initiated
How the company first told customers
Letter

All recalls from CORIN MEDICAL, LTD.

Timeline

FDA's Enforcement Report is its public list of recalls; FDA posts a recall by adding it there.

  1. CORIN MEDICAL began the recall

  2. FDA posted the recall to its Enforcement Report and rated it Class II (moderate risk)

  3. Added to Recall Canary

Summary by Recall Canary, written with AI · summary updated Nov 4, 2026 · primary source: U.S. Food and Drug Administration (opens in a new tab)