MEDICAL DEVICE RECALLLOW

Cardiosave Battery Charging Station

Datascope is recalling Cardiosave Battery Charging Stations due to a screw defect in the left bay that prevents proper battery charging.

What to do

Inspect the left battery bay for a protruding screw. If a screw is present, do not use the left bay for charging; use only the right bay until the unit is repaired or replaced by the manufacturer.

Posted
Updated
Recall No.
Z-1587-2026
Agency
Firm
Datascope Corp.
Recall ID
98432
Status
ONGOING

Why it was recalled

The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in the left battery bay limits full insertion of certain batteries and prevents proper charging. The right bay of the Battery Charging Station is not affected by this issue.

How to identify it

UPC 10607567111964

Battery Charging Station;

Where it was sold

Distributed in US.