Datascope Cardiosave Battery Charging Station
Datascope is recalling the Cardiosave battery charging station due to a screw defect in the left bay that prevents proper charging.
Inspect the left battery bay of your charging station for a protruding screw. If the screw prevents full insertion of the battery, discontinue use of the left bay and contact Getinge for repair or replacement instructions.
Why it was recalled
Datascope is recalling the Cardiosave dual-bay battery charging station because a protruding screw in the left battery bay can prevent full insertion and proper charging of lithium-ion batteries. The right bay of the device is not affected. This recall is distributed nationwide.
Is mine affected?
Check the model, lot, or serial number on the device or its packaging. Yours is affected if it matches one below.
0998-00-0802
- Model
- 0998-00-0802
Where it was sold
Distributed nationwide.
Official source
Other recent recalls from Datascope Corp.
Editorial summary by Recall Canary · last updated 2026-06-11 · primary source: U.S. Food and Drug Administration · severity reflects the agency's published classification