Diagnostic Grifols Erytra and Erytra Eflexis Automated Blood Grouping and Antibody Test Systems

The Erytra blood grouping system has an incorrect configuration that can cause minor variations in titration results.

Recall ended Oct 7, 2026. Still check yours.

What happened

Diagnostic Grifols, S.A. began a recall on March 9, 2026, for the Erytra Eflexis Automated Blood Grouping and Antibody Test System and the Erytra Automated Blood Grouping and Antibody Test System, due to an issue in the technique definition for serial titration of antibodies using distributed DG Gel 8 Neutral cards.

The incorrect configuration, as configured in Templates Pack versions 15.0 and 16.0 for the U.S. market, can lead to minor variations in titration results up to 1 titre, which remains within the generally accepted variability for semi-quantitative titration techniques.

The affected products include the Erytra Eflexis system with Model no. 210600 and the Erytra system with Model nos. 210400 and 210401. The recall was added to the FDA's Enforcement Report and classified on October 7, 2026, and was in the FDA's weekly enforcement report on October 14, 2026. The instruments were updated for three customers in Alabama, Florida and Ohio.

What to do

We don't have product-specific instructions for this recall. FDA's general advice for medicine recalls:

  1. Check the FDA or company notice for instructions specific to this product.
  2. Ask your pharmacist or doctor whether your lot is affected and what to do.

Is yours affected?

2 products

Affected if its UPC is listed.

A UPC is the number printed under a barcode.

  1. Erytra Eflexis Automated Blood Grouping and Antibody Test System, Templates Pack versions 15.0.0 and 16.0.0, Model no. 210600

    UPC
    • 08436583730942
    • 15.0.0
    • 16.0.0
  2. Erytra Automated Blood Grouping and Antibody Test System, Templates Pack versions 15.0.0 and 16.0.0, Model no. 210400, 210401

    UPC
    • 08436583730935
    • 15.0.0
    • 16.0.0

Recall details

Company began the recall
Sold in

There are Erytra and Erytra Eflexis instruments that has been update for three customers through the US. Customers including Alabama, Florida and Ohio.

Company
Diagnostic Grifols, S.A.
Risk

Low risk

Taking it is unlikely to cause health problems.

FDA Class III recall · How risk is rated

Event ID
98557
Recall numbers
B-0267-2026, B-0286-2026

All recalls from Diagnostic Grifols, S.A.

Timeline

FDA's Enforcement Report is its public list of recalls; FDA adds a recall when it posts it and publishes the list weekly.

  1. Company began the recall

  2. Added to FDA's Enforcement Report · Classified by FDA · Recall ended

  3. Listed on Recall Canary

  4. In FDA's weekly enforcement report

Summary by Recall Canary, written with AI · summary updated Oct 12, 2026 · primary source: U.S. Food and Drug Administration (opens in a new tab)