MEDICAL DEVICE RECALLMODERATE

MOSAIQ Oncology Information System

Elekta is recalling the MOSAIQ Oncology Information System due to a software error that may prevent the loading of baseline shift plans.

What to do

Contact your Elekta representative for software updates and follow the manufacturer's provided instructions for verifying treatment plans before delivery.

Posted
Updated
Agency
Firm
Elekta, Inc.
Recall ID
98980
Status
ONGOING

Why it was recalled

The MOSAIQ Oncology Information System is being recalled because the software may fail to load an adapted baseline shift plan during treatment delivery if a communication error occurs. This failure could result in the delivery of incorrect radiation doses to patients.