Skip to main content
Recall Canary
Ferno Power F2 Powered Fastening System
← Back to all recalls
MEDICAL DEVICE RECALLMODERATE

Ferno Power F2 Powered Fastening System

Ferno is recalling the Power F2 Powered Fastening System due to a risk of the rail mechanism lodging and preventing stretcher movement.

What to do

Contact Ferno-Washington for inspection and repair instructions to address the potential roller assembly interference.

Posted
Updated
Agency
Firm
Ferno-Washington Inc
Recall ID
99291
Status
ONGOING

Why it was recalled

Ferno-Washington is recalling the Power F2 Powered Fastening System due to accelerated roller wear and potential deformation. This interference between the roller assembly and rail can cause the mechanism to lodge, preventing the loading or unloading of stretchers during both powered and manual operation.

Is mine affected?

This recall covers the product below. If you have it, yours is affected.

Power F2 Powered Fastening System, Product No. 0567091 Powered fastening system intended for professional use only. Designed to load, unload, and secure a compatible wheeled stretcher inside a ground-based ambulance. Item no. 0567091

← Back to all recalls

Editorial summary by Recall Canary · last updated 2026-07-20 · primary source: U.S. Food and Drug Administration · severity reflects the agency's published classification