FDI Automated Field Steam Sterilizer

FDI recalls Automated Field Steam Sterilizer over possible ground continuity loss in the water recovery system connector.

What to do

Stop using the sterilizer immediately and contact Fort Defiance Industries for instructions on inspection and repair of the 120V Amphenol connector.

Recalled
Listed
Updated
Recall No.
Z-2983-2026
Agency
FDA
Firm
Fort Defiance Industries, LLC
Recall ID
99371
Status
ONGOING

Why it was recalled

Fort Defiance Industries is recalling its Automated Field Steam Sterilizer because the water recovery system's 120V Amphenol connector may intermittently lose ground continuity, creating an unreliable fault current return path.

Is mine affected?

Check the model, lot, or serial number on the device or its packaging. Yours is affected if it matches one below.

Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131

Lot code
All serial numbers/ GTIN: (01) 00862992000327
UDI-DI
00862992000327

Where it was sold

Distributed nationwide.