FoundationOne CDx Sequencing Agent SEQ0067

Foundation Medicine is recalling Sequencing Agent SEQ0067 component of FoundationOne CDx due to possible false-positive BARD1 variants.

What to do

If you have received FoundationOne CDx test results using Sequencing Agent SEQ0067, contact Foundation Medicine immediately to discuss the potential impact on your results and whether retesting is recommended.

Recalled
Listed
Updated
Recall No.
Z-1927-2026
Agency
FDA
Firm
Foundation Medicine, Inc.
Recall ID
98625
Status
ONGOING

Recall at a glance

Quantity recalled
163 units

Why it was recalled

Foundation Medicine is recalling Sequencing Agent SEQ0067, a component of the FoundationOne CDx (F1CDx) genomic testing system, due to a possible increase in phasing that could result in false-positive BARD1 splice site indel artifact variants in test results.