GE Healthcare AW Server 3.2 ext.6.5
GE Healthcare is recalling the AW Server 3.2 ext.6.5 image processing system due to a software error that may display the wrong patient's data.
Contact your GE HealthCare representative or service engineer to schedule a software update to resolve the context synchronization issue. Ensure clinical staff verify patient identity against the displayed images before interpretation or reporting.
Recall at a glance
- Quantity recalled
- 340 servers
Why it was recalled
GE Healthcare is recalling the AW Server 3.2 ext.6.5 radiological image processing system due to a software synchronization error. This issue may cause the system to display the previous patient's exam instead of the intended patient's exam when launching interactive applications, creating a risk of misdiagnosis or incorrect clinical decisions without any system warning.
Is mine affected?
Check the model, lot, or serial number on the device or its packaging. Yours is affected if it matches one below.
GE Healthcare AW Server 3.2 ext.6.5; Radiological Image Processing System
- UDI-DI
- 00840682102384, 00840682102384
Where it was sold
Distributed nationwide.