Portrait Core Services Software V1.1
GE Medical recalls Portrait Core Services Software V1.1 over possible loss of patient monitoring and alarm reset issues.
Contact GE Medical Systems Information Technologies for instructions on updating the software to correct the issue.
Why it was recalled
GE Medical Systems Information Technologies is recalling Portrait Core Services Software V1.1, Ref# 5700209, due to a software issue that may result in a loss of patient monitoring at the Hub and Central Viewer. Additionally, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.
Is mine affected?
Check the model, lot, or serial number on the device or its packaging. Yours is affected if it matches one below.
| Product | Recall # | UDI-DI |
|---|---|---|
| PORTRAIT CORE SERVICES SOFTWARE V1.1, Ref# 5700209 | Z-3036-2026 | 00195278679284 |
| PORTRAIT CORE SERVICES SOFTWARE V1.0.5, Ref# 5697353 | Z-3038-2026 | 00195278364838 |
| FRU (Field replaceable unit), PORTRAIT CORE SERVICES SW RELOAD V1.1, Ref# 5700225 | Z-3039-2026 | 00195278679307 |
| PORTRAIT COMPUTE AND CORE SERVICES V1.2, Ref# 5700592 | Z-3037-2026 | 00198953056752 |
Where it was sold
Distributed in Michigan and New York.
Timeline
- Firm initiated recall
- Publicly announced
- Listed on Recall Canary
- Summary updated
Official source
Other Software Defect recalls
Other recent recalls from GE Medical Systems Information Technologies Inc
All recalls from GE Medical Systems Information Technologies Inc
Editorial summary by Recall Canary · summary updated 2026-08-28 · primary source: U.S. Food and Drug Administration · severity reflects the agency's published classification