FDA recall, , Moderate risk
GE Medical Systems Computed Tomography Systems
GE Medical Systems Computed Tomography Systems may start an X-ray scan without a physical press of the Start Scan button.
What to do
Check yoursIf a code on your device is listed below:
- Check the lot against the recall notice and ask your provider or supplier before you stop using it.
Typical advice for a recalled medical device; the FDA notice gives no product-specific instructions.
What happened
GE Medical Systems, LLC is recalling certain GE Medical Systems Computed Tomography Systems at the request of the U.S. Food and Drug Administration. During use of certain service tools, a sequence of events including scan aborts and pressing the Setup button may allow an X-ray scan to start without a physical press of the Start Scan button.
The system is intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications. The company began the recall on September 21, 2026, and distribution is nationwide. The agency added the recall to its Enforcement Report on November 25, 2026.
Is yours affected?
Match the UDI-DI and serial number listed under your product.
A UDI-DI is the device identifier printed with the barcode on the device label.
The system is intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications
- UDI-DI
- 00198953085950
- Serial number
- PCT8X2600003CN
Recall details
- Posted by FDA
- Recall began
- Status
- Open
- Weekly archive
- Listed the week of Nov 23, 2026
- Sold in
Distributed nationwide and in other countries.
- Recalled by
- GE Medical Systems, LLCWaukesha, Wisconsin
- Hazard
- Device defect
- Risk
Moderate risk
Using it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.
FDA Class II recall · How risk is rated
- FDA event number
- 100233
- Recall number
- Z-0312-2027
- Recall type
- At FDA's request
All recalls from GE Medical Systems, LLC
Timeline
FDA's Enforcement Report is its public list of recalls; FDA posts a recall by adding it there.
GE Medical Systems began the recall
FDA posted the recall to its Enforcement Report and rated it Class II (moderate risk)
Added to Recall Canary