Skip to main content
Sunday email

FDA recall, , Moderate risk

GE Medical Systems Computed Tomography Systems

GE Medical Systems Computed Tomography Systems may start an X-ray scan without a physical press of the Start Scan button.

Product photo

What to do

Check yours

If a code on your device is listed below:

  1. Check the lot against the recall notice and ask your provider or supplier before you stop using it.

Typical advice for a recalled medical device; the FDA notice gives no product-specific instructions.

What happened

GE Medical Systems, LLC is recalling certain GE Medical Systems Computed Tomography Systems at the request of the U.S. Food and Drug Administration. During use of certain service tools, a sequence of events including scan aborts and pressing the Setup button may allow an X-ray scan to start without a physical press of the Start Scan button.

The system is intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications. The company began the recall on September 21, 2026, and distribution is nationwide. The agency added the recall to its Enforcement Report on November 25, 2026.

Is yours affected?

Match the UDI-DI and serial number listed under your product.

A UDI-DI is the device identifier printed with the barcode on the device label.

  1. The system is intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications

    UDI-DI
    • 00198953085950
    Serial number
    • PCT8X2600003CN

Recall details

Posted by FDA
Recall began
Status
Open
Sold in

Distributed nationwide and in other countries.

Recalled by
GE Medical Systems, LLCWaukesha, Wisconsin
Risk

Moderate risk

Using it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.

FDA Class II recall · How risk is rated

FDA event number
100233
Recall number
Z-0312-2027
Recall type
At FDA's request

All recalls from GE Medical Systems, LLC

Timeline

FDA's Enforcement Report is its public list of recalls; FDA posts a recall by adding it there.

  1. GE Medical Systems began the recall

  2. FDA posted the recall to its Enforcement Report and rated it Class II (moderate risk)

  3. Added to Recall Canary

Summary by Recall Canary, written with AI · summary updated Nov 26, 2026 · Official source: U.S. Food and Drug Administration (opens in a new tab) recall · cite: “GE Medical Systems Computed Tomography Systems Recall.” Recall Canary. https://recallcanary.com/recall/fda/ge-medical-systems-the-system-is.