Hamilton Medical Ventilator Coaxial Breathing Circuit
← Back to all recalls
MEDICAL DEVICE RECALLCRITICAL

Hamilton Medical Ventilator Coaxial Breathing Circuit

Hamilton Medical is recalling ventilator coaxial breathing circuits due to a risk of valve membrane adhesion and inadequate ventilation.

What to do

Prior to use, perform a test lung action. If an "exhalation obstructed" alarm occurs, remove the expiratory valve set to overcome membrane adhesion or replace the valve set entirely.

Posted
Updated
Recall No.
Z-2463-2026
Agency
Firm
Hamilton Medical AG
Recall ID
99028
Status
ONGOING

Why it was recalled

Hamilton Medical is recalling coaxial breathing circuit sets used with HAMILTON-C1, HAMILTON-T1, and HAMILTON-MR1 ventilators. The expiratory valve membranes may stick to the valve body, potentially causing an "exhalation obstructed" alarm, inadequate ventilation, and oxygen desaturation. This issue may not be detectable during pre-operative testing.

Is mine affected?

Check the model, lot, or serial number on the device or its packaging. Yours is affected if it matches one below.

Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.

UPC
07630002802956, 17630002802953, 07630002802963, 17630002802960, 07630002802970, 17630002802977, 07630002802987, 17630002802984
Recall #
Z-2463-2026

Where it was sold

Distributed nationwide.

Other recent recalls from Hamilton Medical AG

← Back to all recalls

Editorial summary by Recall Canary · last updated 2026-07-01 · primary source: U.S. Food and Drug Administration · severity reflects the agency's published classification