Skip to main content

FDA recall, , Moderate risk

PALACOS Bone Cement Mixing and Application Systems

Palacos bone cement may have ampoules that do not break, so the product cannot be used on the patient.

What happened

Heraeus Medical GmbH (Dental Division) of Wehrheim, Germany, recalled Palacos MV+G pro 40, Palacos R pro 40, Palacos MV pro 40, and Palacos R+G pro 40 radiopaque poly(methyl methacrylate)-based bone cements, pre-filled into a mixing and application system, due to an increase of complaints regarding breakage of the ampoules within their device systems.

An extensive investigation revealed that multiple factors impact ampoule breakage, and an unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed.

The recall began December 9, 2025, and was added to the FDA's Enforcement Report on January 14, 2026. Distribution was international, including Aland Islands, Australia, Belgium, Denmark, Finland, Germany, Ireland, Italy, Liechtenstein, Netherlands, New Zealand, Norway, Portugal, Spain, Sweden, Switzerland, and the United Kingdom.

What to do

We don't have product-specific instructions for this recall. FDA's general advice for medical device recalls:

  1. A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
  2. Talk to your health care provider before you stop using it.

Is yours affected?

4 products

Affected if its UDI-DI and model/catalog number are listed under that product.

A UDI-DI is the device identifier printed with the barcode on the device label.

  1. PALACOS MV+g Pro 40; PALACOS MV+g Pro 80

    UDI-DI
    • 04260102136243
    • 04260102136274
    Model/Catalog number
    • 5081290
    • 5081291
  2. PALACOS R Pro 40; PALACOS R Pro 80

    UDI-DI
    • 04260102136120
    • 04260102136151
    Model/Catalog number
    • 5081286
    • 5081287
  3. PALACOS MV Pro 40; PALACOS MV Pro 80

    UDI-DI
    • 04260102135338
    • 04260102135369
    Model/Catalog number
    • 5150633
    • 5150635
  4. PALACOS R+g Pro 40; PALACOS R+g Pro 80

    UDI-DI
    • 04260102136182
    • 04260102136212
    Model/Catalog number
    • 5081288
    • 5081289

Recall details

Company began the recall
Sold in

International distribution in the countries of Aland Islands, Australia, Belgium, Denmark, Finland, Germany, Ireland, Italy, Liechtenstein, Netherlands, New Zealand, Norway, Portugal, Spain, Sweden, Switzerland, United Kingdom.

Company
Heraeus Medical GmbH (Dental Division)Wehrheim, Germany
Risk

Moderate risk

Using it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.

FDA Class II recall · How risk is rated

Event ID
98078
Recall numbers
Z-1080-2026 to Z-1083-2026

All recalls from Heraeus Medical GmbH (Dental Division)

Timeline

FDA's Enforcement Report is its public list of recalls; FDA adds a recall when it posts it and publishes the list weekly.

  1. Company began the recall

  2. Added to FDA's Enforcement Report · Classified by FDA

  3. In FDA's weekly enforcement report

  4. Listed on Recall Canary

Summary by Recall Canary, written with AI · summary updated Oct 23, 2026 · primary source: U.S. Food and Drug Administration (opens in a new tab)