MICS3 Angled Sagittal Saw Attachment
Howmedica Osteonics is recalling the MICS3 Angled Sagittal Saw Attachment because external screws may loosen and detach during use.
Immediately stop using the affected MICS3 Angled Sagittal Saw Attachment and contact the manufacturer for further instructions regarding the return or repair of the device.
Why it was recalled
A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.
How to identify it
UPC 07613327523874
MICS3 Angled Sagittal Saw Attachment;
Where it was sold
Distributed in US.
Official source