Hotline Fluid Warming Set
ICU Medical is recalling Hotline Fluid Warming Sets due to a risk of leakage at the tubing-to-connector bond.
Immediately inspect inventory for the affected lots and discontinue use of any identified units. Contact ICU Medical customer service for return and replacement instructions.
Why it was recalled
ICU Medical is recalling specific lots of the Hotline Fluid Warming Set due to a risk of leakage at the tubing-to-connector bond caused by inadequate solvent application. This defect may compromise the delivery of warmed blood and intravenous fluids to patients.
Is mine affected?
This recall covers the product below. If you have it, yours is affected.
Hotline Fluid Warming Set, single use device, is designed to warm blood and intravenous fluids and deliver them to the patient's intravenous access site at normothermic temperatures under gravity flow conditions, List Number L-70
Official source
Editorial summary by Recall Canary · last updated 2026-07-20 · primary source: U.S. Food and Drug Administration · severity reflects the agency's published classification