GEM Premier 5000/7000 Bar Code Reader Wand
← Back to all recalls
MEDICAL DEVICE RECALLMODERATE

GEM Premier 5000/7000 Bar Code Reader Wand

Instrumentation Laboratory recalls BCR Wand for GEM Premier 5000/7000 analyzers; may assign results to wrong patient ID.

What to do

Stop using the affected Bar Code Reader (BCR) Wand immediately. Contact Instrumentation Laboratory for instructions on receiving a corrected configuration update or replacement wand. Do not use the wand to scan patient samples until the issue is resolved.

Recalled
Listed
Updated
Recall No.
Z-2821-2026
Agency
FDA
Firm
Instrumentation Laboratory
Recall ID
99115
Status
ONGOING

Why it was recalled

Instrumentation Laboratory is recalling the Bar Code Reader (BCR) Wand accessory for GEM Premier 5000 and GEM Premier 7000 blood analyzer systems. BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after April 8, 2026) may assign patient results to an incorrect patient ID due to a race condition, potentially leading to inappropriate clinical management.

Is mine affected?

Check the model, lot, or serial number on the device or its packaging. Yours is affected if it matches one below.

ProductRecall #UDI-DIModel/Catalog number
GEM Premier 5000Z-2825-202600024019255, 08426950870227, 25028719, 18021546, 18021549, 26012036, 18051850, 23054764, 21068342, 25080219, 26010781, 25122012, 25041639, 25120558, 25079944, 25079943, 25028719, 25100485, 25079947, 2510048300024019255 Portable blood analyzer system
GEM Premier 5000Z-2823-202600024019255, 08426950870227, 20106073, 20106039, 19052933, 20116415, 20064792, 22062065, 22062067, 19103490, 22092716, 22102881, 18112339, 19012518, 20126582, 20075256, 20064794, 20106036, 24016057, 2108875800024019255 portable blood analyzer system
GEM Premier 5000Z-2824-202600024019255, 08426950870227, 25110547, 25100462, 25110522, 25100472, 25101897, 25101920, 25101931, 25122029, 26012032, 25122030, 26012033, 26012043, 25110523, 25110536, 25120550, 25069638, 23101109, 2401113700024019255 portable blood analyzer system
GEM Premier 7000 with iQM3Z-2822-202600000015279, 0195226000924, 26012062, 26012065, 26012069, 26012070, 00024015859, 0842695086963400000015279 Intelligent Quality Management 3 (iQM3) is used as the quality control (QC) and assessment system for the GEM Premier 7000. portable blood analyzer
GEM Premier 5000 with IQM2 Model/Catalog Number: 00024019255 and GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Portable blood analyzer system. Affected Accessory: Bar Code Reader (BCR) Wand Part No. 00024015859Z-2821-202600024019255, 08426950870227, 00000015279, 0195226000924, 00024015859, 08426950869634
GEM Premier 5000Z-2826-202600024019255, 08426950870227, 25100466, 24097933, 23064897, 00024015859, 0842695086963400024019255 Portable blood analyzer system
GEM Premier 500000024019255
GEM Premier 7000 with iQM30195226000924, 7000, 0842695086963400000015279 Intelligent Quality Management 3 (iQM3) is used as the quality control (QC) and assessment system for the GEM Premier 7000. portable blood analyzer
GEM Premier 5000 with IQM2 Model/Catalog Number: 00024019255 and GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Portable blood analyzer system. Affected Accessory: Bar Code Reader (BCR) Wand Part No. 0002401585908426950870227, 7000, 0195226000924, 2026, 08426950869634

Where it was sold

Distributed nationwide.

Timeline

  1. Firm initiated recall
  2. Publicly announced
  3. Listed on Recall Canary
  4. Summary updated

Other Device Defect recalls

Other recent recalls from Instrumentation Laboratory

All recalls from Instrumentation Laboratory