FDA recall, , Moderate risk
Instrumentation Laboratory GEM Premier 5000 and GEM Premier 7000 Blood Analyzers
GEM Premier 5000 and 7000 systems with BCR wands may assign results to wrong patient.
What happened
Instrumentation Laboratory of Bedford, Massachusetts, recalled GEM Premier 5000 and GEM Premier 7000 portable blood analyzer systems with Bar Code Reader (BCR) wands because the wands may assign patient results to an incorrect patient ID. The BCR wands received a configuration update released under Service Information No. 2026-07 on or after April 8, 2026. The root cause is a race condition introduced when the "price check" feature was disabled to reduce excess current draw.
The recall, begun May 8, 2026, covers US Nationwide distribution. Affected products include the GEM Premier 5000 with IQM2 and the GEM Premier 7000 with iQM3, both using the BCR Wand Part No. 00024015859. The FDA added the recall to its Enforcement Report on July 29, 2026.
2 photos
What to do
We don't have product-specific instructions for this recall. FDA's general advice for medical device recalls:
- A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
- Talk to your health care provider before you stop using it.
Is yours affected?
Compare the details below with your product and check the FDA notice to confirm whether it is affected.
A UDI-DI is the device identifier printed with the barcode on the device label.
GEM Premier 5000 with IQM2 Model/
Product details from the notice
GEM Premier 5000 with IQM2 Model/Catalog Number: 00024019255 and GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Portable blood analyzer system. Affected Accessory: Bar Code Reader (BCR) Wand Part No. 00024015859
- UDI-DI
- 00024019255
- 08426950870227
- 00000015279
- 0195226000924
- 00024015859
- 08426950869634
GEM Premier 5000
- UDI-DI
- 00024019255
- 08426950870227
- 25028719
- 18021546
- 18021549
- 26012036
- 18051850
- 23054764
- 21068342
- 25080219
- 26010781
- 25122012
- 25041639
- 25120558
- 25079944
- 25079943
- 25100485
- 25079947
- 25100483
- 25100484
- 21057870
- 26010710
- 00024015859
- 08426950869634
- Model/Catalog number
- 00024019255 Portable blood analyzer system
- UDI-DI
- 00024019255
- 08426950870227
- 20106073
- 20106039
- 19052933
- 20116415
- 20064792
- 22062065
- 22062067
- 19103490
- 22092716
- 22102881
- 18112339
- 19012518
- 20126582
- 20075256
- 20064794
- 20106036
- 24016057
- 21088758
- 21088759
- 18061910
- 24016081
- 24016108
- 24108036
- 24108044
- 24067370
- 24067371
- 24077468
- 00024015859
- 08426950869634
- Model/Catalog number
- 00024019255 portable blood analyzer system
- UDI-DI
- 00024019255
- 08426950870227
- 25110547
- 25100462
- 25110522
- 25100472
- 25101897
- 25101920
- 25101931
- 25122029
- 26012032
- 25122030
- 26012033
- 26012043
- 25110523
- 25110536
- 25120550
- 25069638
- 23101109
- 24011137
- 25081799
- 25081797
- 25091839
- 25091840
- 25091864
- 25091870
- 25091868
- 25091873
- 25091880
- 25091876
- 25101886
- 25101889
- 25101887
- 25101891
- 25101907
- 25101904
- 25101912
- 25101905
- 25101917
- 25101921
- 25101926
- 25101929
- 25101934
- 25101927
- 25111968
- 25121986
- 25121985
- 25121981
- 25121984
- 25121994
- 25121995
- 25121982
- 25121993
- 25121991
- 25121999
- 25121996
- 16030255
- 17060782
- 17060784
- 17060785
- 17060789
- 17060790
- 17060797
- 17111126
- 17121258
- 18011414
- 20054515
- 20054516
- 22082271
- 00024015859
- 08426950869634
- Model/Catalog number
- 00024019255 portable blood analyzer system
- UDI-DI
- 00024019255
- 08426950870227
- 25100466
- 24097933
- 23064897
- 00024015859
- 08426950869634
- Model/Catalog number
- 00024019255 Portable blood analyzer system
GEM Premier 7000 with iQM3
- UDI-DI
- 00000015279
- 0195226000924
- 26012062
- 26012065
- 26012069
- 26012070
- 00024015859
- 08426950869634
- Model/Catalog number
- 00000015279 Intelligent Quality Management 3 (iQM3) is used as the quality control (QC) and assessment system for the GEM Premier 7000. portable blood analyzer
Recall details
- Company began the recall
- Sold in
US Nationwide distribution.
- Company
- Instrumentation LaboratoryBedford, Massachusetts
- Hazard
- Device defect
- Risk
Moderate risk
Using it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.
FDA Class II recall · How risk is rated
- Event ID
- 99115
- Recall numbers
- Z-2821-2026 to Z-2826-2026
All recalls from Instrumentation Laboratory
Timeline
FDA's Enforcement Report is its public list of recalls; FDA adds a recall when it posts it and publishes the list weekly.
Company began the recall
Added to FDA's Enforcement Report · Classified by FDA
Listed on Recall Canary
In FDA's weekly enforcement report
Summary updated