
Da Vinci Surgical System E-Brake Retainer Pin
Intuitive Surgical is recalling the Da Vinci surgical system E-brake retainer pin due to a risk of uncontrolled motion and patient injury.
Contact Intuitive Surgical or your local representative to schedule a service inspection and replacement of the affected E-brake retainer pin.
Why it was recalled
Intuitive Surgical is recalling the E-brake retainer pin used in certain Da Vinci surgical systems. A defect in the pin may prevent the E-brake from engaging or cause it to back out, potentially leading to uncontrolled motion of the surgical column and a risk of patient tissue injury, crushing, or bleeding.
Is mine affected?
Check the model, lot, or serial number on the device or its packaging. Yours is affected if it matches one below.
Da Vinci ASSY, PSS, SP1098, Part Number: 380601
- UPC
- 00886874114605, 10008062, 10211920, 10003059, 10009706, 10003509, 10003510, 10009707, 10065786, 10020729, 10012963, 10020705, 10031145, 10031144, 10012962, 10036960, 10711066, 10035476, 10043635, 10072852
- Recall #
- Z-2521-2026
Where it was sold
Distributed nationwide.
Official source
Other recent recalls from Intuitive Surgical, Inc.
Editorial summary by Recall Canary · last updated 2026-06-23 · primary source: U.S. Food and Drug Administration · severity reflects the agency's published classification
