Skip to main content
Recall Canary
KayserBett IDA Pediatric Care Bed
← Back to all recalls
MEDICAL DEVICE RECALLCRITICAL

KayserBett IDA Pediatric Care Bed

KayserBett is recalling the IDA pediatric care bed due to an entrapment risk if the hand control is not locked while the child is unattended.

What to do

Ensure the hand control adjustment functions are locked with the provided key whenever the patient is unattended to prevent unauthorized operation.

Recalled
Listed
Updated
Recall No.
Z-2212-2026
Agency
FDA
Firm
KAYSERBETTEN GMBH & CO. KG
Recall ID
98820
Status
ONGOING

Recall at a glance

Quantity recalled
372 units

Why it was recalled

KayserBett is recalling the IDA pediatric care bed due to a risk of entrapment and injury. If the hand control adjustment functions are not locked with the provided key while the child is unattended, the patient or others may operate the controls, leading to potential entrapment under the bed frame or between the frame and the floor.

Is mine affected?

This recall covers the product below. If you have it, yours is affected.

Pediatric care bed; Product Designation: KayserBett IDA;

Where it was sold

Distributed nationwide.

Timeline

  1. Firm initiated recall
  2. Publicly announced
  3. Listed on Recall Canary
  4. Summary updated
← Back to all recalls

Editorial summary by Recall Canary · summary updated 2026-06-11 · primary source: U.S. Food and Drug Administration · severity reflects the agency's published classification