Skip to main content

FDA recall, , High risk

Medline Semi Electric and Full Electric Homecare Beds

Medline beds have a potential hazard involving the hand control pendant and wires, which may overheat and pose a fire risk.

What happened

Medline Industries, LP, of Northfield, Illinois, has identified a potential hazard involving the hand control pendant and associated wires on several models of Medline beds. In rare and atypical scenarios, such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components or motor, the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire.

The company began the recall on November 26, 2025, and the FDA added it to its Enforcement Report on January 16, 2026. The affected products include Semi Electric Basic Homecare Bed, Full Electric Basic Homecare Bed, Semi Electric Lightweight Homecare Bed, Full Electric Lightweight Homecare Bed, Full Electric Low Lightweight Homecare Bed, and Full Electric Low Basic Homecare Bed. The beds were distributed in the US, including California, Pennsylvania, Guam, and VI.

What to do

We don't have product-specific instructions for this recall. FDA's general advice for medical device recalls:

  1. A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
  2. Talk to your health care provider before you stop using it.

Is yours affected?

6 products

Compare the details below with your product and check the FDA notice to confirm whether it is affected.

A UDI-DI is the device identifier printed with the barcode on the device label.

  1. MEDLINE beds labeled as: 1) Semi Electric Basic Homecare Bed, REF MDR107002E; 2) Semi Electric Basic Homecare Bed, 4-pack,REF MDR107002E-4

    UDI-DI
    • 40080196320999
    Lot code
    All lots; 2) UDI/DI 40080196320999, All Lots
  2. MEDLINE beds labeled as: 1) Full Electric Basic Homecare Bed, REF MDR107003E; 2) Full Electric Basic Homecare Bed, 4-pack,REF MDR107003E-4.

    UDI-DI
    • 40080196321002
    Lot code
    All lots; 2) REF MDR107003E-4, UDI/DI 40080196321002, All lots
  3. MEDLINE beds labeled as: Semi Electric Lightweight Homecare Bed, REF MDR107002L

    UDI-DI
    • 40080196294320
    Lot code
    All lots
  4. MEDLINE beds labeled as: Full Electric Lightweight Homecare Bed, REF MDR107003L

    UDI-DI
    • 40080196294344
    Lot code
    All lots
  5. MEDLINE beds labeled as: Full Electric Low Lightweight Homecare Bed, REF MDR107003LO

    UDI-DI
    • 40080196325901
    Lot code
    All lots
  6. MEDLINE beds labeled as: Full Electric Low Basic Homecare Bed, REF MDR107003ELO

    UDI-DI
    • 10080196546992
    Lot code
    All lots

Recall details

Company began the recall
Company
Medline Industries, LPNorthfield, Illinois
Risk

High risk

There is a reasonable chance that using it could cause serious health problems or death.

FDA Class I recall · How risk is rated

Event ID
98185
Recall numbers
Z-1011-2026 to Z-1016-2026

All recalls from Medline Industries, LP

Timeline

FDA's Enforcement Report is its public list of recalls; FDA adds a recall when it posts it and publishes the list weekly.

  1. Company began the recall

  2. Added to FDA's Enforcement Report · Classified by FDA

  3. In FDA's weekly enforcement report

  4. Listed on Recall Canary

Summary by Recall Canary, written with AI · summary updated Oct 23, 2026 · primary source: U.S. Food and Drug Administration (opens in a new tab)