Medtronic MiniMed 780G Insulin Pump

A software issue may cause Pump Error 43 or 41, suspending insulin delivery.

What happened

Medtronic MiniMed, Inc. of Northridge, California is recalling 577 units of the MiniMed 780G Insulin Pump with Software Version 6.60. A software timing and processor communication issue may trigger Pump Error 43 and/or Pump Error 41, resulting in suspension of insulin delivery. The pump is indicated for patients age 7-80 years with Type 1 diabetes.

The company began the recall on November 2, 2025, and the FDA added it to the Enforcement Report on February 2, 2026. The product was distributed worldwide, including U.S. national distribution across many states and the country of Israel.

What to do

We don't have product-specific instructions for this recall. FDA's general advice for medical device recalls:

  1. A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
  2. Talk to your health care provider before you stop using it.

Is yours affected?

Match the list below, then check the codes the FDA notice also names.

A UDI-DI is the device identifier printed with the barcode on the device label.

  1. MiniMed 780G Insulin Pump Catalog Numbers

    UDI-DI
    • 00199150003686
    • 00763000414344
    • 00763000545604
    • 00763000578640
    • 00763000586195
    • 00763000640095
    • 00763000758547
    • 00763000819378
    • 00763000819385
    • 00763000819392
    • 00763000865665
    • 00763000924089
    • 00763000924102
    • 00763000924119
    • 76300075854701
    • 76300081937801
    • 76300081938501
    • 76300096001801
    • 76300096003201
    • 76300096004901
    • 76300062988501

From the FDA notice

MiniMed 780G Insulin Pump Catalog Numbers: MMT-1884 MMT-1886 With Software Version 6.60 The MiniMed 780G insulin pump is indicated for use by patients age 7-80 years with Type 1 diabetes, whose total daily dose of insulin is 8 units per day or more.

Recall details

Company began the recall
Company
Medtronic MiniMed, Inc.Northridge, California
Risk

Moderate risk

Using it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.

FDA Class II recall · How risk is rated

Event ID
97997
Recall number
Z-1238-2026
Quantity recalled
577 units

All recalls from Medtronic MiniMed, Inc.

Timeline

FDA's Enforcement Report is its public list of recalls; FDA adds a recall when it posts it and publishes the list weekly.

  1. Company began the recall

  2. Added to FDA's Enforcement Report · Classified by FDA

  3. In FDA's weekly enforcement report

  4. Listed on Recall Canary

Summary by Recall Canary, written with AI · summary updated Oct 23, 2026 · primary source: U.S. Food and Drug Administration (opens in a new tab)