Octopus Nuvo Tissue Stabilizer
← Back to all recalls
MEDICAL DEVICE RECALLMODERATE

Octopus Nuvo Tissue Stabilizer

Medtronic is recalling Octopus Nuvo Tissue Stabilizer (TSMICS1) due to incorrect tubing assembly that may affect surgical tissue stabilization.

What to do

Stop using the affected Octopus Nuvo Tissue Stabilizer units immediately. Contact Medtronic Perfusion Systems for instructions on product return or replacement. Verify the Product Number TSMICS1 on your device.

Posted
Updated
Recall No.
Z-1473-2026
Agency
Firm
Medtronic Perfusion Systems
Recall ID
98354
Status
ONGOING

Why it was recalled

Medtronic is recalling the Octopus Nuvo Tissue Stabilizer (Product Number TSMICS1), a disposable tissue stabilization system used during cardiac surgery, due to an assembly error where tubing was incorrectly positioned—the short tube was attached to the bottom of the canister and the long tube to the top, contrary to the intended design. This misalignment may affect proper tissue stabilization during surgical procedures. Affected units were distributed nationwide in Florida and Indiana.

Is mine affected?

Check the model, lot, or serial number on the device or its packaging. Yours is affected if it matches one below.

Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product Number TSMICS1

UPC
00763000543693, 0232837110, 0231547054, 0231591827, 0231758880, 0231758898, 0231758932, 0231792414, 0231792421, 0231912903, 0232670901, 0232709219, 0232709254, 0232744928, 0232784925, 0232784989, 0232805458
Recall #
Z-1473-2026

Where it was sold

Distributed nationwide.