Merge PACS Medical Imaging Workstation
Merge PACS workstation recalled over risk of displaying thumbnails from a prior patient study.
Contact Merge Healthcare to determine if your workstation is affected and to receive instructions for updating the software to correct this issue.
Why it was recalled
Merge Healthcare is recalling its Merge PACS medical imaging diagnostic workstation because thumbnails from a previously opened patient study may be displayed in the Thumbnail Bar while a different patient study is open, potentially leading to misidentification of patient images.
Is mine affected?
Check the model, lot, or serial number on the device or its packaging. Yours is affected if it matches one below.
| Product | Recall # | UDI-DI |
|---|---|---|
| Merge PACS, medical imaging diagnostic workstation | Z-2874-2026 | 00842000101215, 00842000101154 |
| Merge OrthoPACS, medical imaging diagnostic workstation | Z-2875-2026 | 00842000101154 |
Where it was sold
Distributed nationwide.
Timeline
- Firm initiated recall
- Publicly announced
- Listed on Recall Canary
- Summary updated
Official source
Other Device Defect recalls
Other recent recalls from Merge Healthcare, Inc.
Editorial summary by Recall Canary · summary updated 2026-09-03 · primary source: U.S. Food and Drug Administration · severity reflects the agency's published classification