Olympus Toomey Evacuator 30cc
← Back to all recalls
MEDICAL DEVICE RECALLMODERATE

Olympus Toomey Evacuator 30cc

Olympus recalls Toomey Evacuator 30cc (Model 188) over sterilization failure risk.

What to do

Stop using the recalled Toomey Evacuator 30cc with Adaptor, Model 188, and contact Olympus Corporation of the Americas for return instructions and further guidance.

Recalled
Listed
Updated
Recall No.
Z-2718-2026
Agency
FDA
Firm
Olympus Corporation of the Americas
Recall ID
98869
Status
ONGOING

Why it was recalled

Olympus is recalling the Toomey Evacuator 30cc with Adaptor, Model 188, because some sterilization parameters in the instructions for use may not consistently render the devices adequately sterile, posing a risk of infection. Consumers should stop using the device and contact Olympus for instructions on returning the affected units.

Is mine affected?

Check the model, lot, or serial number on the device or its packaging. Yours is affected if it matches one below.

ProductModel
Olympus Toomey Evacuator 30cc with Adaptor.188
Olympus Toomey Evacuator 50cc with Adaptor.188A
Olympus Ellik Evacuator Includes Glass Body, Male Connecting Tubing & Latex Bulb.194
Olympus Glass Body for 194 Ellik Evacuator.190A

Where it was sold

Distributed nationwide.

Timeline

  1. Firm initiated recall
  2. Publicly announced
  3. Listed on Recall Canary
  4. Summary updated