Skip to main content

FDA recall, , Moderate risk

Ortho-Clinical Diagnostics ORTHO VISION and ORTHO VISION Max Analyzers

Quidel Ortho is recalling the Ortho Vision and Ortho Vision Max Analyzers for ID-Micro Typing System (MTS) Gel Cards Software Version 5.16 and below for a software anomaly where the analyzer is unable to perform inventory of loaded reagents if the loading station door is opened immediately after the allotted 20-second countdown ends but within 0.2 seconds before to the door re-locks.

Recall ended Oct 13, 2026. Still check yours.

What happened

Quidel Ortho is recalling the ORTHO VISION Analyzer for ID-MTS Gel Cards (Product Code 6904577) and the ORTHO VISION Max Analyzer for ID-MTS Gel Cards (Product Code 6904576), Software Version 5.16 and below, for a software anomaly where the analyzer is unable to perform inventory of loaded reagents if the loading station door is opened immediately after the allotted 20-second countdown ends but within 0.2 seconds before the door re-locks.

The recall began on April 22, 2025, and was added to the FDA's Enforcement Report on March 25, 2026. The affected analyzers were distributed to 50 states, the District of Columbia, Canada, Colombia, Japan, Singapore, the United Kingdom, and Germany.

2 photos

What to do

We don't have product-specific instructions for this recall. FDA's general advice for medicine recalls:

  1. Check the FDA or company notice for instructions specific to this product.
  2. Ask your pharmacist or doctor whether your lot is affected and what to do.

Is yours affected?

3 products

Compare these details with yours, then check the recall notice to see which units are recalled.

  1. Ortho Vision Max Analyzer for ID-MTS Gel Cards (Ortho Vision Max), Product Code 6904576, Version 5.16

  2. Ortho Vision Analyzer for ID-MTS Gel Cards (Ortho Vision), Product Code 6904577, Version 5.16

  3. Ortho Vision Analyzer for ID-MTS Gel Cards (Certified), Product Code 6153434, Version 5.16

    Affected units
    not provided; not distributed

Recall details

Company began the recall
Company
Ortho-Clinical Diagnostics, Inc.
Risk

Moderate risk

Taking it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.

FDA Class II recall · How risk is rated

Event ID
96842
Recall numbers
B-0030-2026, B-0041-2026, B-0042-2026

Timeline

FDA's Enforcement Report is its public list of recalls; FDA adds a recall when it posts it and publishes the list weekly.

  1. Company began the recall

  2. Added to FDA's Enforcement Report · Classified by FDA

  3. In FDA's weekly enforcement report

  4. Recall ended

  5. Listed on Recall Canary