Skip to main content

FDA recall, , Moderate risk

Permobil M5 Corpus Power Wheelchair

The M5 Corpus Power Wheelchair may drive under restricted conditions due to missing drive inhibit logic.

What happened

This recall is being reported retroactively. Power Platform firmware is missing specific drive inhibit logic. This issue may allow the chair to drive under restricted conditions, and may create an unstable condition in certain circumstances.

What to do

We don't have product-specific instructions for this recall. FDA's general advice for medical device recalls:

  1. A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
  2. Talk to your health care provider before you stop using it.

Is yours affected?

Affected if its UDI-DI matches.

A UDI-DI is the device identifier printed with the barcode on the device label.

  1. M5 Corpus Power Wheelchair REF,120032

    UDI-DI
    • 17330818334562

Recall details

Company began the recall
Company
Permobil IncLebanon, Tennessee
Risk

Moderate risk

Using it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.

FDA Class II recall · How risk is rated

Event ID
99267
Recall number
Z-2805-2026
Quantity recalled
715 wheelchairs

All recalls from Permobil Inc

Timeline

FDA's Enforcement Report is its public list of recalls; FDA adds a recall when it posts it and publishes the list weekly.

  1. Company began the recall

  2. Added to FDA's Enforcement Report · Classified by FDA

  3. In FDA's weekly enforcement report

  4. Listed on Recall Canary