Skip to main content

FDA recall, , High risk

Philips Lumify Diagnostic Ultrasound System

The Philips Lumify Diagnostic Ultrasound System may not recognize transducer registration, prevent scanning, or have unstable connections.

What happened

Philips Ultrasound, LLC recalled the Philips Lumify Diagnostic Ultrasound System because it may not recognize previous transducer registration, prevent scanning, and require re-registration with Wi-Fi/cellular access, which may be unavailable outside the hospital, or have an unstable connection at the transducer USB cable to tablet/phone that may disconnect. The company began the recall on September 24, 2026, and the FDA added it to the Enforcement Report on October 16, 2026.

What to do

We don't have product-specific instructions for this recall. FDA's general advice for medical device recalls:

  1. A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
  2. Talk to your health care provider before you stop using it.

Is yours affected?

Affected if its UDI-DI is listed.

A UDI-DI is the device identifier printed with the barcode on the device label.

  1. Philips Lumify Diagnostic Ultrasound System

    UDI-DI
    • 00884838122116
    • 1.X
    • 3.X
    • 4.X
    • 5.X
    • 00884838124141
    • 2.X

Recall details

Company began the recall
Company
Philips Ultrasound, LLC
Risk

High risk

There is a reasonable chance that using it could cause serious health problems or death.

FDA Class I recall · How risk is rated

Event ID
99878

All recalls from Philips Ultrasound, LLC

Timeline

FDA's Enforcement Report is its public list of recalls; FDA adds a recall when it posts it and publishes the list weekly.

  1. Company began the recall

  2. Added to FDA's Enforcement Report · Classified by FDA

  3. Listed on Recall Canary

Summary by Recall Canary, written with AI · summary updated Oct 17, 2026 · primary source: U.S. Food and Drug Administration (opens in a new tab)