Philips Spectral CT on Rails
Philips Spectral CT on Rails (Model 728334) requires a software update to address gantry positioning, message display, and menu option issues.
Contact Philips Medical Systems immediately to obtain and install the required software update for your Spectral CT on Rails system (Model 728334). Do not perform affected scan procedures until the update is applied.
Recall at a glance
- Quantity recalled
- 3 systems
Why it was recalled
Philips Spectral CT on Rails (Model 728334) is subject to a software update due to three identified issues: the gantry may remain at the current scan position after pressing Go during continuous CT scans; a required message prompt for previous surview selection may not display; and the Define Head Area option may be grayed out after setting Brain Area DoseRight Index. Contact Philips Medical Systems for the software update.
Is mine affected?
Check the model, lot, or serial number on the device or its packaging. Yours is affected if it matches one below.
Philips Spectral CT on Rails.
- Model
- 728334
Where it was sold
Distributed nationwide.