Skip to main content

FDA recall, , Moderate risk

RaySearch Laboratories RayStation/RayPlan Radiation Therapy Treatment Planning Systems

RayStation/RayPlan radiation therapy treatment planning system may not invalidate calculated radiation dose for certain regions of interest.

What happened

RAYSEARCH LABORATORIES AB, of Stockholm, Sweden, is recalling RayStation/RayPlan radiation therapy treatment planning systems. The potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs).

This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

What to do

We don't have product-specific instructions for this recall. FDA's general advice for medical device recalls:

  1. A device recall doesn't always mean you must stop using it: it may need to be checked, adjusted or fixed.
  2. Talk to your health care provider before you stop using it.

Is yours affected?

Compare the details below with your product and check the FDA notice to confirm whether it is affected.

A UDI-DI is the device identifier printed with the barcode on the device label.

  1. Product details

    RayStation/RayPlan. Model/Catalog Numbers: 12.0.0, 12.1.0, 12.0.3, 12.0.4, 12.3.0. Software Version: RayStation 11B, RayStation 11B SP1, RayStation 11B SP2, RayStation 11B SP3, RayStation 11B SPT1. Radiation Therapy Treatment Planning System

    UDI-DI
    • 07350002010426
    • 07350002010495
    • 07350002010501
    • 07350002010600
    • 07350002010570
  2. RayStation/RayPlan

    1. UDI-DI
      • 07350002010778
      Model/Catalog number
      • 16.0.0
      Software version
      RayStation 2024B
    2. UDI-DI
      • 07350002010549
      • 07350002010679
      • 07350002010730
      Model/Catalog number
      • 13.0.0.1547
      • 13.1.0.144
      • 13.1.1.89
      Software version
      RayStation 12A, RayStation 12A SP1, RayStation 12A SP2
    3. UDI-DI
      • 07350002010471
      Model/Catalog number
      • 10.1.1
      Software version
      RayStation 10B SP1
    4. UDI-DI
      • 07350002010822
      • 07350002010945
      Model/Catalog number
      • 17.0.0
      • 17.0.1
      Software version
      RayStation v2025, RayStation v2025 SP1
    5. UDI-DI
      • 07350002010761
      • 07350002010808
      Model/Catalog number
      • 15.2.0
      • 15.1.3
      Software version
      RayStation 2024A SP1, RayStation 2024A SP2
    6. UDI-DI
      • 07350002010556
      • 07350002010723
      • 07350002010884
      Model/Catalog number
      • 14.0.0
      • 15.0.0
      • 15.1.3
      Software version
      RayStation 2023B, RayStation 2024A, RayStation 2024A SP3
  3. RayStation/RayPlan. Model/Catalog Number: 10.1. Software Version: RayStation/RayPlan 10B. Radiation Therapy Treatment Planning System

    UDI-DI
    • 07350002010310
  4. RayStation/RayPlan. Model/Catalog Numbers: 11.0.0, 11.0.1, 11.0.3, 11.0.4. Software Version: RayStation 11A, RayStation 11A SP1, RayStation 11A SP2, RayStation 11A SP3. Radiation Therapy Treatment Planning System

    UDI-DI
    • 07350002010389
    • 07350002010433
    • 07350002010440
    • 07350002010631

Recall details

Company began the recall
Sold in

Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS

Company
RAYSEARCH LABORATORIES ABStockholm, Sweden
Risk

Moderate risk

Using it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.

FDA Class II recall · How risk is rated

Event ID
98117
Recall numbers
Z-1102-2026 to Z-1110-2026

All recalls from RAYSEARCH LABORATORIES AB

Timeline

FDA's Enforcement Report is its public list of recalls; FDA adds a recall when it posts it and publishes the list weekly.

  1. Company began the recall

  2. Added to FDA's Enforcement Report · Classified by FDA

  3. In FDA's weekly enforcement report

  4. Listed on Recall Canary