FDA recall, , Moderate risk
RaySearch Laboratories RayStation Radiation Therapy Treatment Planning Systems
RayStation planning software may cause dose deformation when a deformable registration expands outwards near the original dose grid boundary.
What to do
Check yoursIf a code on your device is listed below:
- Check the lot against the recall notice and ask your provider or supplier before you stop using it.
Typical advice for a recalled medical device; the FDA notice gives no product-specific instructions.
What happened
RaySearch Laboratories AB, of Stockholm, Sweden, is recalling its RayStation radiation therapy treatment planning systems. The company started this recall on its own. The software may have a potential for dose deformation when a deformable registration expands outwards near the boundary of the original dose grid. This possibly causes the clinically relevant dose to be lost.
Affected models include RayStation 2023B and RayStation 2024A SP3, among others. The systems were distributed worldwide, with nationwide distribution in the United States. The company began the recall on October 7, 2026, and told customers by letter. The FDA added the recall to its Enforcement Report on November 18, 2026.
Is yours affected?
Compare the details below with your product and check the FDA notice to confirm whether it is affected.
A UDI-DI is the device identifier printed with the barcode on the device label.
RayStation Systems: RayStation 6.1.1, RayStation 6 SP2
Product details from the notice
RayStation Systems: RayStation 6.1.1, RayStation 6 SP2. Model/Build Numbers: 6.1.1.2, 6.2.0.7. Software Radiation Therapy Treatment Planning System
- UDI-DI
- 07350002010082
- 07350002010075
- Other code
- 0735000201008220170529
- 0735000201007520170630
- Serial number
- 6.1.1.2
- 6.2.0.7
RayStation 2023B
Product details from the notice
RayStation 2023B. Model/Build Number: 14.0.0.3338. Radiation Therapy Treatment Planning System
- UDI-DI
- 07350002010556
- Other code
- 0735000201055620230630
- Serial number
- 14.0.0.3338
RayStation Systems: RayStation 9B SP1, RayStation 10A SP1
Product details from the notice
RayStation Systems: RayStation 9B SP1, RayStation 10A SP1. Model/Build Numbers: 9.2.0.483, 10.0.1.52. Radiation Therapy Treatment Planning System
- UDI-DI
- 07350002010297
- 07350002010365
- Other code
- 0735000201029720200310
- 0735000201036520200526
- Serial number
- 9.2.0.483
- 10.0.1.52
RayStation Systems: RayStation 12A, RayStation 12A SP1
Product details from the notice
RayStation Systems: RayStation 12A, RayStation 12A SP1. Model/Build Numbers: 13.0.0.1547, 13.1.0.144. Radiation Therapy Treatment Planning System
- UDI-DI
- 07350002010549
- 07350002010679
- Other code
- 0735000201054920220616
- 0735000201067920221007
- Serial number
- 13.0.0.1547
- 13.1.0.144
RayStation 10B
Product details from the notice
RayStation 10B. Model/Build Number: 10.1.0.613 Radiation Therapy Treatment Planning System
- UDI-DI
- 07350002010310
- Other code
- 0735000201031020201216
- Serial number
- 10.1.0.613
RayStation Systems: RayStation 11A, RayStation 11A SP1, RayStation 11A SP2
Product details from the notice
RayStation Systems: RayStation 11A, RayStation 11A SP1, RayStation 11A SP2. Model/Build Numbers: 11.0.0.951, 11.0.1.29, 11.0.3.116 Radiation Therapy Treatment Planning System
- UDI-DI
- 07350002010389
- 07350002010433
- 07350002010440
- Other code
- 0735000201038920210518
- 0735000201043320210610
- 0735000201044020210916
- Serial number
- 11.0.0.951
- 11.0.1.29
- 11.0.3.116
RayStation Systems: RayStation 7, RayStation 8A, RayStation 8A SP1, RayStation 8B SP1, RayStation 8B SP2, RayStation 9A
Product details from the notice
RayStation Systems: RayStation 7, RayStation 8A, RayStation 8A SP1, RayStation 8B SP1, RayStation 8B SP2, RayStation 9A. Model/Build Numbers: 7.0.0.19, 8.0.0.61, 8.0.1.10, 8.1.1.8, 8.1.2.5, 9.0.0.113. Radiation Therapy Treatment Planning System
- UDI-DI
- 07350002010068
- 07350002010112
- 07350002010136
- 07350002010204
- 07350002010235
- 07350002010174
- Other code
- 0735000201006820171130
- 0735000201011220180608
- 0735000201013620180918
- 0735000201020420190214
- 0735000201023520190524
- 0735000201017420190612
- Serial number
- 7.0.0.19
- 8.0.0.61
- 8.0.1.10
- 8.1.1.8
- 8.1.2.5
- 9.0.0.113
RayStation 2024A SP3
Product details from the notice
RayStation 2024A SP3. Model/Build Number: 15.1.3.10. Radiation Therapy Treatment Planning System
- UDI-DI
- 07350002010884
- Other code
- 0735000201088420250129
- Serial number
- 15.1.3.10
RayStation Systems: RayStation 11B, RayStation 11B SP2, RayStation 11B SP3, RayStation 11B SP1, RayStation 11B SPT1
Product details from the notice
RayStation Systems: RayStation 11B, RayStation 11B SP2, RayStation 11B SP3, RayStation 11B SP1, RayStation 11B SPT1. Model/Build Numbers: 12.0.0.932, 12.0.3.68, 12.0.4.12, 12.1.0.1221, 12.3.0.119. Radiation Therapy Treatment Planning System
- UDI-DI
- 07350002010495
- 07350002010426
- 07350002010501
- 07350002010600
- 0735000201057
- Other code
- 0735000201042620211208
- 0735000201050120220422
- 0735000201060020220620
- 0735000201057020221222
- Serial number
- 12.0.0.932
- 12.0.3.68
- 12.0.4.12
- 12.1.0.1221
- 12.3.0.119
Recall details
- Posted by FDA
- Recall began
- Status
- Open
- Weekly archive
- Listed the week of Nov 16, 2026
- Sold in
Distributed nationwide and in other countries.
- Recalled by
- Raysearch Laboratories ABStockholm, Sweden
- Hazard
- Device defect
- Risk
Moderate risk
Using it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.
FDA Class II recall · How risk is rated
- FDA event number
- 100093
- Recall numbers
- Z-0483-2027 to Z-0491-2027
- Recall type
- Started by the company
- How the company first told customers
- Letter
All recalls from Raysearch Laboratories AB
Timeline
FDA's Enforcement Report is its public list of recalls; FDA posts a recall by adding it there.
RAYSEARCH LABORATORIES AB began the recall
FDA posted the recall to its Enforcement Report and rated it Class II (moderate risk)
Added to Recall Canary