Skip to main content

FDA recall, , Moderate risk

RaySearch Laboratories RayStation Radiation Therapy Treatment Planning Systems

RayStation planning software may cause dose deformation when a deformable registration expands outwards near the original dose grid boundary.

What to do

Check yours

If a code on your device is listed below:

  1. Check the lot against the recall notice and ask your provider or supplier before you stop using it.

Typical advice for a recalled medical device; the FDA notice gives no product-specific instructions.

What happened

RaySearch Laboratories AB, of Stockholm, Sweden, is recalling its RayStation radiation therapy treatment planning systems. The company started this recall on its own. The software may have a potential for dose deformation when a deformable registration expands outwards near the boundary of the original dose grid. This possibly causes the clinically relevant dose to be lost.

Affected models include RayStation 2023B and RayStation 2024A SP3, among others. The systems were distributed worldwide, with nationwide distribution in the United States. The company began the recall on October 7, 2026, and told customers by letter. The FDA added the recall to its Enforcement Report on November 18, 2026.

Is yours affected?

Compare the details below with your product and check the FDA notice to confirm whether it is affected.

A UDI-DI is the device identifier printed with the barcode on the device label.

  1. RayStation Systems: RayStation 6.1.1, RayStation 6 SP2

    Product details from the notice

    RayStation Systems: RayStation 6.1.1, RayStation 6 SP2. Model/Build Numbers: 6.1.1.2, 6.2.0.7. Software Radiation Therapy Treatment Planning System

    UDI-DI
    • 07350002010082
    • 07350002010075
    Other code
    • 0735000201008220170529
    • 0735000201007520170630
    Serial number
    • 6.1.1.2
    • 6.2.0.7
  2. RayStation 2023B

    Product details from the notice

    RayStation 2023B. Model/Build Number: 14.0.0.3338. Radiation Therapy Treatment Planning System

    UDI-DI
    • 07350002010556
    Other code
    • 0735000201055620230630
    Serial number
    • 14.0.0.3338
  3. RayStation Systems: RayStation 9B SP1, RayStation 10A SP1

    Product details from the notice

    RayStation Systems: RayStation 9B SP1, RayStation 10A SP1. Model/Build Numbers: 9.2.0.483, 10.0.1.52. Radiation Therapy Treatment Planning System

    UDI-DI
    • 07350002010297
    • 07350002010365
    Other code
    • 0735000201029720200310
    • 0735000201036520200526
    Serial number
    • 9.2.0.483
    • 10.0.1.52
  4. RayStation Systems: RayStation 12A, RayStation 12A SP1

    Product details from the notice

    RayStation Systems: RayStation 12A, RayStation 12A SP1. Model/Build Numbers: 13.0.0.1547, 13.1.0.144. Radiation Therapy Treatment Planning System

    UDI-DI
    • 07350002010549
    • 07350002010679
    Other code
    • 0735000201054920220616
    • 0735000201067920221007
    Serial number
    • 13.0.0.1547
    • 13.1.0.144
  5. RayStation 10B

    Product details from the notice

    RayStation 10B. Model/Build Number: 10.1.0.613 Radiation Therapy Treatment Planning System

    UDI-DI
    • 07350002010310
    Other code
    • 0735000201031020201216
    Serial number
    • 10.1.0.613
  6. RayStation Systems: RayStation 11A, RayStation 11A SP1, RayStation 11A SP2

    Product details from the notice

    RayStation Systems: RayStation 11A, RayStation 11A SP1, RayStation 11A SP2. Model/Build Numbers: 11.0.0.951, 11.0.1.29, 11.0.3.116 Radiation Therapy Treatment Planning System

    UDI-DI
    • 07350002010389
    • 07350002010433
    • 07350002010440
    Other code
    • 0735000201038920210518
    • 0735000201043320210610
    • 0735000201044020210916
    Serial number
    • 11.0.0.951
    • 11.0.1.29
    • 11.0.3.116
  7. RayStation Systems: RayStation 7, RayStation 8A, RayStation 8A SP1, RayStation 8B SP1, RayStation 8B SP2, RayStation 9A

    Product details from the notice

    RayStation Systems: RayStation 7, RayStation 8A, RayStation 8A SP1, RayStation 8B SP1, RayStation 8B SP2, RayStation 9A. Model/Build Numbers: 7.0.0.19, 8.0.0.61, 8.0.1.10, 8.1.1.8, 8.1.2.5, 9.0.0.113. Radiation Therapy Treatment Planning System

    UDI-DI
    • 07350002010068
    • 07350002010112
    • 07350002010136
    • 07350002010204
    • 07350002010235
    • 07350002010174
    Other code
    • 0735000201006820171130
    • 0735000201011220180608
    • 0735000201013620180918
    • 0735000201020420190214
    • 0735000201023520190524
    • 0735000201017420190612
    Serial number
    • 7.0.0.19
    • 8.0.0.61
    • 8.0.1.10
    • 8.1.1.8
    • 8.1.2.5
    • 9.0.0.113
  8. RayStation 2024A SP3

    Product details from the notice

    RayStation 2024A SP3. Model/Build Number: 15.1.3.10. Radiation Therapy Treatment Planning System

    UDI-DI
    • 07350002010884
    Other code
    • 0735000201088420250129
    Serial number
    • 15.1.3.10
  9. RayStation Systems: RayStation 11B, RayStation 11B SP2, RayStation 11B SP3, RayStation 11B SP1, RayStation 11B SPT1

    Product details from the notice

    RayStation Systems: RayStation 11B, RayStation 11B SP2, RayStation 11B SP3, RayStation 11B SP1, RayStation 11B SPT1. Model/Build Numbers: 12.0.0.932, 12.0.3.68, 12.0.4.12, 12.1.0.1221, 12.3.0.119. Radiation Therapy Treatment Planning System

    UDI-DI
    • 07350002010495
    • 07350002010426
    • 07350002010501
    • 07350002010600
    • 0735000201057
    Other code
    • 0735000201042620211208
    • 0735000201050120220422
    • 0735000201060020220620
    • 0735000201057020221222
    Serial number
    • 12.0.0.932
    • 12.0.3.68
    • 12.0.4.12
    • 12.1.0.1221
    • 12.3.0.119

Recall details

Posted by FDA
Recall began
Status
Open
Sold in

Distributed nationwide and in other countries.

Recalled by
Raysearch Laboratories ABStockholm, Sweden
Risk

Moderate risk

Using it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.

FDA Class II recall · How risk is rated

FDA event number
100093
Recall numbers
Z-0483-2027 to Z-0491-2027
Recall type
Started by the company
How the company first told customers
Letter

All recalls from Raysearch Laboratories AB

Timeline

FDA's Enforcement Report is its public list of recalls; FDA posts a recall by adding it there.

  1. RAYSEARCH LABORATORIES AB began the recall

  2. FDA posted the recall to its Enforcement Report and rated it Class II (moderate risk)

  3. Added to Recall Canary

Summary by Recall Canary, written with AI · summary updated Nov 20, 2026 · Official source: U.S. Food and Drug Administration (opens in a new tab) recall · cite: “RaySearch Laboratories RayStation Radiation Therapy Treatment Planning Systems Recall.” Recall Canary. https://recallcanary.com/recall/fda/raysearch-raystation-systems-raystation-6-1-1-raystation.