Shimadzu Medical Systems X-Ray and Angiography Generators
← Back to all recalls
MEDICAL DEVICE RECALLMODERATE

Shimadzu Medical Systems X-Ray and Angiography Generators

Shimadzu is recalling 20+ models of X-ray and angiography systems due to a contactor defect that may keep power active when switched off.

What to do

Contact your local Shimadzu service representative to schedule a mandatory inspection and repair of the high-voltage generator contactors.

Posted
Updated
Agency
Firm
Shimadzu Medical Systems
Recall ID
98571
Status
ONGOING

Why it was recalled

Electromagnetic contactors within high voltage generator may weld together, resulting in current being supplied to power supply even if power is off.

How to identify it

Digital Angiography System Trinias:UD150G-40 / D150GC-40;

← Back to all recalls

Editorial summary by Recall Canary · last updated 2026-06-03 · primary source: U.S. Food and Drug Administration · severity reflects the agency's published classification