FDA recall, , Moderate risk

Siemens Healthineers Atellica CH HDL Cholesterol and Total Protein_2 Tests

Atellica CH HDL Cholesterol and Total Protein_2 reagents may cause falsely elevated or depressed test results.

What happened

Siemens Healthcare Diagnostics, Inc. is recalling two test products, Atellica CH Total Protein_2 and Atellica CH HDL Cholesterol. The company says reagent probe carryover may cause falsely elevated or falsely depressed results. Carryover from HDL Cholesterol may cause falsely elevated Hemoglobin A1c results. Carryover from Total Protein_2 may cause falsely depressed Direct Bilirubin 2 results. These results may affect patient samples, quality control, and calibrator results.

The company started the recall on its own and told customers by letter, beginning September 22, 2026. The recall covers 25210 units of Atellica CH Total Protein_2 and 72207 units of Atellica CH HDL Cholesterol. Distribution is nationwide in the United States and worldwide in other countries. The agency classified the recall on November 6, 2026.

What to do

If a code on your device is listed below:

  1. Check the lot against the recall notice and ask your provider or supplier before you stop using it.

Typical advice for a recalled medical device; the FDA notice gives no product-specific instructions.

Is yours affected?

2 products

Compare the details below with your product and check the FDA notice to confirm whether it is affected.

A UDI-DI is the device identifier printed with the barcode on the device label.

  1. Atellica CH Total Protein_2 (TP_2)

    Product details from the notice

    Atellica CH Total Protein_2 (TP_2);
    Siemens Material Number (SMN): 1. 11645755 (3700 tests), 2. 11537228 (7400 tests);

    UDI-DI
    • 00630414654010
    • 00630414611082
    Lot code
    All lots; 2. Siemens Material Number (SMN): 11537228 (7400 tests); UDI-DI: 00630414611082; All lots;
  2. Atellica CH HDL Cholesterol (HDLC)

    Product details from the notice

    Atellica CH HDL Cholesterol (HDLC);
    Siemens Material Number (SMN): 11537213;

    UDI-DI
    • 00630414610832
    Lot code
    All lots;

Recall details

Posted by FDA
Recall began
Status
Open
Sold in

Distributed nationwide and in Argentina, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Burkina Faso, Canada, Chile, China, Colombia, Costa Rica, Croatia, Curaçao, St Eus, Czech Republic, Democratic Republic Congo, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Maldives, Mexico, Morocco, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Turkmenistan, U.A.E, Uganda, United Kingdom, Uruguay, Uzbekistan, Vatikancity and Vietnam.

Recalled by
Siemens Healthcare Diagnostics, Inc.Tarrytown, New York
Risk

Moderate risk

Using it may cause short-term health problems, or ones that treatment can reverse. Serious harm is unlikely.

FDA Class II recall · How risk is rated

FDA event number
100035
Recall numbers
Z-0364-2027 to Z-0365-2027
Recall type
Started by the company
How the company first told customers
Letter

All recalls from Siemens Healthcare Diagnostics, Inc.

Timeline

FDA's Enforcement Report is its public list of recalls; FDA posts a recall by adding it there.

  1. Siemens Healthcare Diagnostics, Inc. began the recall

  2. FDA posted the recall to its Enforcement Report and rated it Class II (moderate risk)

  3. Added to Recall Canary

Summary by Recall Canary, written with AI · summary updated Nov 7, 2026 · primary source: U.S. Food and Drug Administration (opens in a new tab) recall · cite: “Siemens Healthineers Atellica CH HDL Cholesterol and Total Protein_2 Tests Recall.” Recall Canary. https://recallcanary.com/recall/fda/siemens-healthcare-diagnostics-atellica-ch-hdl-cholesterol-hdlc.