EVO+ Visian Toric Implantable Collamer Lens (TICL)

Staar Surgical is recalling EVO+ Visian Toric ICL Model VTICM5 due to incorrect length labeling (12.1 mm labeled vs. 12.6 mm actual) causing excessive vault and potential angle closure.

What to do

If you have received or implanted an EVO+ Visian Toric Implantable Collamer Lens Model VTICM5, contact Staar Surgical AG immediately to verify the actual lens length and discuss whether lens exchange or removal is necessary. Do not delay seeking medical evaluation if you experience eye discomfort, vision changes, or symptoms of elevated intraocular pressure.

Posted
Updated
Recall No.
Z-1480-2026
Agency
Firm
Staar Surgical AG
Recall ID
98397
Status
ONGOING

Why it was recalled

Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.

How to identify it

UPC 00840311304363

EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5