COAPTIUM Connect
TISSIUM recalls COAPTIUM Connect nerve repair device over unsubstantiated 36-month shelf life.
Check your inventory for the COAPTIUM Connect device and discontinue use of units beyond the revised 26-month shelf life. Contact TISSIUM SA for relabeling instructions and further guidance.
Why it was recalled
TISSIUM SA is recalling the COAPTIUM Connect, an in situ polymerizing peripheral nerve repair device, because the currently labeled 36-month shelf life is not substantiated by real-time data. Devices will be relabeled to reflect a revised shelf life of 26 months. Use of the device beyond its validated shelf life may result in perioperative breakage, rendering the device unusable and potentially requiring revision surgery.
Is mine affected?
Check the model, lot, or serial number on the device or its packaging. Yours is affected if it matches one below.
| Product | Lot code | Recall # | UDI-DI | Model/Catalog number |
|---|---|---|---|---|
| COAPTIUM CONNECT | NCOCO2010, 250700042, 251000033, 251000034 | Z-2903-2026 | 03760279380094, 250700042, 251000033, 251000034 | NCOCO2010 |
| COAPTIUM CONNECT | NCOCO4010, 251000041, 250700048, 260300046, 260400050, 260300047 | Z-2906-2026 | 03760279380148, 251000041, 250700048, 260300046, 260400050, 260300047 | NCOCO4010 |
| COAPTIUM CONNECT | NCOCO5010, 250700050, 251000043, 260300048, 260300049, 260300050 | Z-2907-2026 | 03760279380179, 250700050, 251000043, 260300048, 260300049, 260300050 | NCOCO5010 |
| COAPTIUM CONNECT | NCOCO2075, 250700016, 251000048, 251000049, 260300043 | Z-2904-2026 | 03760279380100, 250700016, 251000048, 251000049, 260300043 | NCOCO2075 |
| COAPTIUM Connect | NCOCO1510, 250700041, 251000032 | Z-2901-2026 | 03760279380070, 250700041, 251000032 | NCOCO1510 |
| COAPTIUM CONNECT | NCOCO1575, 250700007, 251000031, 260300041 | Z-2902-2026 | 03760279380087, 250700007, 251000031, 260300041 | NCOCO1575 |
| COAPTIUM CONNECT | NCOCO6010, 250700051, 251000044, 251000045 | Z-2908-2026 | 03760279380209, 250700051, 251000044, 251000045 | NCOCO6010 |
| COAPTIUM CONNECT | NCOCO3010, 250700045, 251000040, 251000036, 251000037, 251000039, 251000035, 251000064 | Z-2905-2026 | 03760279380117, 250700045, 251000040, 251000036, 251000037, 251000039, 251000035, 251000064 | NCOCO3010 |
Where it was sold
Distributed nationwide.